Abstract

An 85-year-old man underwent abdominal aortic aneurysm (AAA) repair with a synthetic graft at age 48. Sac enlargement, likely due to an endoleak, was observed during follow-up, and covered stents were placed in the graft limbs at age 80. Despite this, the sac continued to enlarge, leading to a referral to our institution. Imaging revealed an endoleak from an uncovered graft section. An EXCLUDER® stent graft (W. L. Gore & Associates, Inc., Flagstaff, AZ) was placed to cover the graft entirely, resolving the endoleak. The patient was discharged on postoperative day 5. This case highlights the rare failure of a synthetic graft after 37 years.

摘要

一名85岁男性在48岁时接受了人工血管修复腹主动脉瘤手术。随访期间发现瘤体增大,可能是由于内漏所致,80岁时在人工血管分支处放置了覆膜支架。尽管如此,瘤体仍继续增大,遂转诊至我院。影像学检查显示未覆盖的人工血管段存在内漏。放置了EXCLUDER®覆膜支架(美国戈尔公司,亚利桑那州弗拉格斯塔夫)以完全覆盖人工血管,解决了内漏问题。患者术后第5天出院。该病例凸显了人工血管在37年后罕见的失败情况。