Suppr超能文献

使用磁共振波谱法测量聚乙烯吡咯烷酮(PVP)溶液中的温度。

Measuring temperature in polyvinylpyrrolidone (PVP) solutions using MR spectroscopy.

作者信息

Gai Neville D, Dong Ruifeng, van der Veen Jan Willem, Ouwerkerk Ronald, Pierpaoli Carlo

机构信息

Laboratory on Quantitative Medical Imaging, NIBIB/NIH, Bethesda, Maryland, USA.

Magnetic Resonance Spectroscopy Core, NIMH/NIH, Bethesda, Maryland, USA.

出版信息

Med Phys. 2025 May;52(5):3151-3160. doi: 10.1002/mp.17683. Epub 2025 Feb 17.

Abstract

BACKGROUND

Polyvinylpyrrolidone (PVP) water solutions could be used for cross-site and cross-vendor validation of diffusion-related measurements. However, since water diffusivity varies as a function of temperature, knowing the temperature of the PVP solution at the time of the measurement is fundamental in accomplishing this task.

PURPOSE

MR spectroscopy (MRS) could provide absolute temperature measurements since the water peak moves relative to any stable peak as temperature changes. In this work, the PVP proton spectrum was investigated to see if any stable peaks would allow for temperature determination. Reproducibility and repeatability for three scanners from three vendors were also assessed.

METHODS

A spherical 17 cm container filled with 40% PVP w/w in distilled water was used for the experiments. A Point REsolved Spectroscopy Sequence (PRESS) with water suppression was employed on three 3T scanners from different vendors-GE, Siemens, and Philips. Frequency separation (in ppm) between peaks was measured in a voxel at the location of a fiber optic temperature probe and mapped to probe measured temperature. The center peak of the first methylene proton triplet closest to water peak was selected for analysis in jMRUI due to its ease of identification and echo time shift invariance. Shift in ppm of the central methylene peak proton was mapped against measured temperatures. Repeatability and reproducibility across the three scanners were determined at room temperature using 10 repeated PRESS scans. MRS established ppm shift versus temperature relationship was used to predict temperature in different PVP phantoms which were then compared against fiber optic probe measured temperature values.

RESULTS

Several H peaks were identified on all scans of the PVP phantom. The water peak moved by ∼-0.01 ppm/°C on the three scanners relative to a central methylene peak. The maximum mean absolute temperature difference over a temperature range of 18-35°C between the three scanners was 0.16°C while the minimum was 0.057°C. Repeatability on each scanner was excellent (std range: 0.00-0.14°C) over 10 repeated PRESS scans. Reproducibility across the three scanners was also excellent with mean temperature difference between scanners ranging between 0.1 and 0.4°C. Temperature values from MRS were within prediction bounds on the three scanners for another in-house prepared 40% PVP phantom (maximum difference<0.3°C), while they were consistently overestimated for another 30% PVP phantom (<1°C) and underestimated for a CaliberMRI 40% PVP phantom (<2.8°C).

CONCLUSIONS

PVP solutions exhibit stable proton peaks, one of which was used for assessing the temperature of the solution using MR proton spectroscopy. These measurements are fast and feasible with standard sequences and postprocessing MRS software and provide fundamental information for calibration of diffusion MRI using PVP solutions.

摘要

背景

聚乙烯吡咯烷酮(PVP)水溶液可用于扩散相关测量的跨站点和跨供应商验证。然而,由于水的扩散率随温度变化,在测量时了解PVP溶液的温度是完成此任务的基础。

目的

磁共振波谱(MRS)可提供绝对温度测量,因为随着温度变化,水峰相对于任何稳定峰移动。在这项工作中,研究了PVP质子谱,以查看是否有任何稳定峰可用于温度测定。还评估了来自三个供应商的三台扫描仪的可重复性和重复性。

方法

使用一个装满40%(w/w)PVP蒸馏水的17厘米球形容器进行实验。在来自不同供应商(GE、西门子和飞利浦)的三台3T扫描仪上采用具有水抑制功能的点分辨波谱序列(PRESS)。在光纤温度探头位置的体素中测量峰之间的频率分离(以ppm为单位),并将其映射到探头测量的温度。由于其易于识别且回波时间偏移不变性,在jMRUI中选择最接近水峰的第一个亚甲基质子三重峰的中心峰进行分析。将中心亚甲基峰质子的ppm偏移与测量温度进行映射。在室温下使用10次重复的PRESS扫描确定三台扫描仪之间的重复性和再现性。利用MRS建立的ppm偏移与温度的关系来预测不同PVP体模中的温度,然后将其与光纤探头测量的温度值进行比较。

结果

在PVP体模的所有扫描中都识别出了几个H峰。在三台扫描仪上,水峰相对于中心亚甲基峰以约-0.01 ppm/°C的速度移动。在18-35°C的温度范围内,三台扫描仪之间的最大平均绝对温度差为0.16°C,最小为0.057°C。在10次重复的PRESS扫描中,每台扫描仪的重复性都非常好(标准差范围:0.00-0.14°C)。三台扫描仪之间的再现性也非常好,扫描仪之间的平均温度差在0.1至0.4°C之间。对于另一个内部制备的40%PVP体模,MRS得出的温度值在三台扫描仪的预测范围内(最大差值<0.3°C),而对于另一个30%PVP体模则始终高估(<1°C),对于CaliberMRI 40%PVP体模则低估(<2.8°C)。

结论

PVP溶液表现出稳定的质子峰,其中一个用于通过MR质子波谱评估溶液温度。这些测量使用标准序列和后处理MRS软件快速可行,并为使用PVP溶液校准扩散MRI提供了基础信息。

相似文献

5
Frequency drift in MR spectroscopy at 3T.3T 磁共振波谱中的频率漂移。
Neuroimage. 2021 Nov 1;241:118430. doi: 10.1016/j.neuroimage.2021.118430. Epub 2021 Jul 24.

本文引用的文献

5
MRI-visible liquid crystal thermometer.磁共振成像可见液晶温度计。
Magn Reson Med. 2020 Sep;84(3):1552-1563. doi: 10.1002/mrm.28224. Epub 2020 Feb 28.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验