Matvijenko K, Borusevičius R
Zalgiris Clinic, Institute of Odontology, Faculty of Medicine, Vilnius University, Vilnius, Lithuania.
Zalgiris Clinic, Institute of Odontology, Faculty of Medicine, Vilnius University, Vilnius, Lithuania.
Int J Oral Maxillofac Surg. 2025 Jul;54(7):647-656. doi: 10.1016/j.ijom.2025.02.005. Epub 2025 Feb 19.
Dynamic navigation is an innovative technology in implant surgery that enhances the precision of implant placement through real-time guidance for clinicians. This technology allows on-the-spot adjustments during surgery, reducing the risk of complications and improving implant outcomes. The aim of this systematic review was to assess the accuracy of various dynamic navigation systems in implant placement using in vitro models. A comprehensive literature search was performed across several databases, focusing on studies published between 2016 and 2024 that reported three-dimensional (3D) and angular deviations. Seven in vitro studies were included, analysing five dynamic navigation systems (ImplaNav, Navident, Denacam, X-Guide, and DCARER), with 649 implants evaluated. Results showed mean coronal 3D deviations between 0.46 mm and 1.58 mm, while apical deviations ranged from 0.48 mm to 2.12 mm. Angular deviations varied between 1.01° and 4.24°. Maximum deviations reached up to 4.80 mm for coronal 3D deviation and 10.70° for angular deviation. All systems demonstrated high accuracy within clinically acceptable limits, with X-Guide showing the lowest numerical errors. Factors like tracking technology, calibration methods, and user experience were found to influence accuracy. Overall, dynamic navigation significantly improves implant placement accuracy compared to freehand methods but remains dependent on technical factors.
动态导航是植入手术中的一项创新技术,它通过为临床医生提供实时引导来提高植入物放置的精度。这项技术允许在手术过程中进行现场调整,降低并发症风险并改善植入效果。本系统评价的目的是使用体外模型评估各种动态导航系统在植入物放置中的准确性。在多个数据库中进行了全面的文献检索,重点关注2016年至2024年期间发表的报告三维(3D)和角度偏差的研究。纳入了七项体外研究,分析了五种动态导航系统(ImplaNav、Navident、Denacam、X-Guide和DCARER),共评估了649颗植入物。结果显示,冠状面3D偏差的平均值在0.46毫米至1.58毫米之间,根尖偏差在0.48毫米至2.12毫米之间。角度偏差在1.01°至4.24°之间变化。冠状面3D偏差的最大偏差达到4.80毫米,角度偏差达到10.70°。所有系统在临床可接受的范围内都显示出高精度,X-Guide的数值误差最低。发现跟踪技术、校准方法和用户体验等因素会影响准确性。总体而言,与徒手方法相比,动态导航显著提高了植入物放置的准确性,但仍依赖于技术因素。