Suppr超能文献

用于结合研究的自动差示热滴定和电位滴定仪。

An automated differential thermal and potentiometric titration apparatus for binding studies.

作者信息

Berger R L, Cascio H E, Davids N, Gibson C G, Marini M, Thiebault L

出版信息

J Biochem Biophys Methods. 1985 Mar;10(5-6):245-59. doi: 10.1016/0165-022x(85)90059-4.

Abstract

A differential pH-thermal titration apparatus is described which can detect pH differences with a sensitivity of +/- 0.0001 pH units and a thermal sensitivity of +/- 0.00002 degree C at a time constant of 0.1 s. With a reaction which yields 1 kcal mol-1, the current system can detect concentrations as low as 4 X 10(-6) M or, in a 2 ml volume, a total amount of 40 nmol. With a time constant of 0.1 s, the sensitivity is 20 +/- 4 micro degrees C. The experimental protocol is specified by a microprocessor and three modes of operation are possible: titration at constant rate of reagent addition, titration at variable rates of addition so that the contents of both cells are at either constant pH or at a constant temperature and variable rate when a rate of change is specified. Experimental data are collected in files, corrected for heat loss, initial baseline drift, and changes in volume. The final corrected data from the standardized run of 0.01338 M HCl in 0.2 M KCl at 25 degrees C calibrate the pH scale and yield the calorimetric conversion constants and pKw which are calculated and stored for subsequent corrections for the titration of an unknown acid or the measurement of binding constants and heats.

摘要

本文描述了一种差分pH - 热滴定仪,它能够检测pH差异,灵敏度为±0.0001个pH单位,热灵敏度为±0.00002℃,时间常数为0.1 s。对于产生1 kcal mol-1热量的反应,当前系统能够检测低至4×10(-6) M的浓度,或者在2 ml体积中检测到总量为40 nmol的物质。在时间常数为0.1 s时,灵敏度为20±4微摄氏度。实验方案由微处理器指定,有三种操作模式:以恒定的试剂添加速率进行滴定、以可变的添加速率进行滴定,以使两个样品池中的物质处于恒定pH或恒定温度,以及在指定变化速率时以可变速率进行滴定。实验数据收集在文件中,并针对热损失、初始基线漂移和体积变化进行校正。在25℃下,0.2 M KCl中0.01338 M HCl的标准化运行的最终校正数据校准了pH标度,并得出量热转换常数和pKw,这些数据经过计算并存储,以便随后对未知酸的滴定或结合常数及热量的测量进行校正。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验