Prashar Sarika Sharma, Kumar Ajay, Sood Ashwani, Mittal Bhagwant Rai
Department of Nuclear Medicine and PET, Postgraduate Institute of Medical Education and Research, Chandigarh, India.
Indian J Nucl Med. 2024 Sep-Oct;39(5):411-414. doi: 10.4103/ijnm.ijnm_81_24. Epub 2025 Jan 25.
Generally after the accreditation of the gamma camera, its unfailing optimal operation is guaranteed with established departmental preventive maintenance schedules. However, it is not easy to always religiously follow the same maintenance routines. Therefore, instead of preventive maintenance schedule, a reactive maintenance schedule was adopted in our department for a short span of time. In our center, the humidity level in the gamma camera room was recorded to be 90%, which was much higher than the recommended value of 50%-60%. This was due to the inundated water in the gamma camera room due to an accidental breakage of fire-extinguishing water pipeline, adjacent to the room. Postaccident, higher values of nonuniformity (both intrinsic and extrinsic uniformity) were observed. The reactive maintenance was undertaken by medical physicist and field engineer to record and confirm the crystal hydration in the detector and to restore the system for the clinical studies. The digital workflow program successfully rectified the problem of crystal hydration. It is highly recommended to perform timely reactive action which may prevent the permanent damage to the systems.
一般来说,在伽马相机获得认证后,通过既定的科室预防性维护计划可确保其始终保持最佳运行状态。然而,始终严格遵循相同的维护程序并非易事。因此,我们科室在短时间内采用了反应性维护计划,而非预防性维护计划。在我们中心,伽马相机室的湿度记录为90%,远高于推荐的50%-60%的值。这是由于与该房间相邻的灭火水管意外破裂,导致伽马相机室被水淹没。事故发生后,观察到不均匀性(固有均匀性和外在均匀性)的值较高。医学物理学家和现场工程师进行了反应性维护,以记录和确认探测器中的晶体水合情况,并恢复系统以进行临床研究。数字工作流程程序成功纠正了晶体水合问题。强烈建议及时采取反应性措施,这可能会防止系统受到永久性损坏。