Suppr超能文献

即时专利磁性十二指肠-回肠吻合术(IMPA-DI):首例人体研究。

Immediately-Patent Magnetic Duodeno-Ileal Anastomosis (IMPA-DI): The First-in-Human Study.

作者信息

Bhandari Mohit, Neto Manoel Galvao, Brunaldi Vitor Ottoboni, Bhandari Mahak, Mathor Winnie, Reddy Manoj, Teixeira Andre, Wilson Erik B

机构信息

Mohak Hitech Speciality Hospital, Indore, India.

Orlando Health, Orlando, USA.

出版信息

Obes Surg. 2025 Jun;35(6):2053-2058. doi: 10.1007/s11695-025-07823-7. Epub 2025 Mar 27.

Abstract

BACKGROUND

Duodeno-ileal anastomoses are technically challenging and associated with high complication rates, limiting their adoption in bariatric surgery despite their efficacy in weight loss. Magnetic compression anastomoses (MCAs) have emerged as a promising technique, promoting tissue fusion and reducing risks of bleeding and leakage. However, delayed patency limits their application. The OTOLoc™ system, combined with Flexagon self-forming magnets (SFMs), addresses this limitation by enabling immediate patency during healing. This study reports the first-in-human experience of immediately patent magnetic duodeno-ileal anastomosis (IMPA-DI) using these novel devices during single anastomosis duodeno-ileal bypass with sleeve gastrectomy (SADI-S).

METHODS

This single-center, observational study evaluated the feasibility and short-term safety of IMPA-DI. Seven adult patients with moderate to severe obesity were included. The SFM assembly, coupled with the OTOLoc™ system, was laparoscopically deployed to create an immediately patent duodeno-ileal anastomosis. Primary endpoints were technical success and freedom from anastomosis-related adverse events (AEs) within 30 days. Secondary outcomes included procedural metrics and overall safety.

RESULTS

All seven procedures were successfully completed, with a median overall duration of 58 min and magnet placement duration of 12 min. Three intraoperative AEs (serosal and liver injuries, staple line bleeding) were unrelated to IMPA and resolved without complications. All patients resumed liquid diets within 2 days, and no anastomosis-related AEs were observed within 30 days. All devices were naturally expelled without complications.

CONCLUSIONS

This first-in-human study demonstrates the feasibility, safety, and efficiency of the IMPA-DI approach using Flexagon SFMs and the OTOLoc™ system. The immediate patency and straightforward deployment highlight the potential for broader application in challenging anastomotic sites. While limited by the small sample size and single-center design, the results underscore the promise of magnetic compression anastomoses. Further studies are warranted to validate these findings and explore long-term outcomes.

摘要

背景

十二指肠-回肠吻合术在技术上具有挑战性,且并发症发生率高,尽管其在减重方面有效,但限制了其在减肥手术中的应用。磁性压缩吻合术(MCA)已成为一种有前景的技术,可促进组织融合并降低出血和渗漏风险。然而,延迟通畅限制了其应用。OTOLoc™系统与Flexagon自成型磁体(SFM)相结合,通过在愈合过程中实现即时通畅来解决这一限制。本研究报告了在单吻合十二指肠-回肠旁路联合袖状胃切除术(SADI-S)期间使用这些新型装置进行即时通畅磁性十二指肠-回肠吻合术(IMPA-DI)的首例人体经验。

方法

这项单中心观察性研究评估了IMPA-DI的可行性和短期安全性。纳入了7名中度至重度肥胖的成年患者。通过腹腔镜部署与OTOLoc™系统相结合的SFM组件,以创建即时通畅的十二指肠-回肠吻合术。主要终点是技术成功和30天内无吻合口相关不良事件(AE)。次要结果包括手术指标和总体安全性。

结果

所有7例手术均成功完成,总中位时长为58分钟,磁体放置时长为12分钟。3例术中AE(浆膜和肝脏损伤、吻合钉线出血)与IMPA无关,且无需并发症即可解决。所有患者在2天内恢复流食,30天内未观察到吻合口相关AE。所有装置均自然排出,无并发症。

结论

这项首例人体研究证明了使用Flexagon SFM和OTOLoc™系统的IMPA-DI方法的可行性、安全性和有效性。即时通畅和简单的部署突出了其在具有挑战性的吻合部位更广泛应用的潜力。虽然受样本量小和单中心设计的限制,但结果强调了磁性压缩吻合术的前景。有必要进行进一步研究以验证这些发现并探索长期结果。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验