Hyasat Bahi, Hani Amjad Bani, Saraireh Ali Al, Kirmeen Rana Al, Sabha Dina, Yamin Saif, Massad Islam, Hammoudeh Ayman
Department of Cardiac Surgery, Queen Alia Heart Institute, Royal Medical Services, Amman, 11855, Jordan.
School of Medicine, University of Jordan, Amman, Jordan.
J Cardiothorac Surg. 2025 Apr 2;20(1):176. doi: 10.1186/s13019-025-03409-7.
Continuous blood gas monitoring (CBGM) during cardiopulmonary bypass (CPB) is essential for maintaining optimal patient outcomes, enabling rapid responses to critical fluctuations in blood gas parameters. This non-inferiority study evaluates the Quantum Perfusion System by Spectrum Medical, which features continuous online blood gas monitoring through Quantum workstation (QWS) and Quantum ventilation module (QVM) without the use of cuvettes, against the standard blood gas analysis (BGA) analyzer to assess real-time clinical accuracy.
This retrospective study included a sample of 40 patients, monitored continuously with the QPS and compared at intervals against standard BGA measurements. The patients undergoing on elective CPB procedures, specifically for coronary artery bypass grafting (CABG), mitral valve replacement (MVR), and aortic valve replacement (AVR).
Pre-alignment deviations for all parameters were within CLIA thresholds, confirming baseline reliability. For hemoglobin, the pre-alignment deviation was 1.9%, which decreased to 0.7% post-alignment, both within the CLIA threshold of ± 5%, with a Bland-Altman mean difference of 0.0988 g/dL (limits: 0.0963 to 0.1012 g/dL). Hematocrit showed a pre-alignment deviation of 2.1%, reduced to 0.2% post-alignment, both within the CLIA threshold of ± 5%, with a Bland-Altman mean difference of 0.3009% (limits: 0.2956 to 0.3063%). For PaO₂, the pre-alignment deviation was 3.9%, reduced to 0.4% post-alignment, both within the CLIA threshold of ± 10%, with a Bland-Altman mean difference of 4.0490 mmHg (limits: 3.9976 to 4.1004 mmHg). PCO₂ demonstrated a pre-alignment deviation of 4.2%, reduced to 0.19% post-alignment, both within the CLIA threshold of ± 10%, with a Bland-Altman mean difference of 0.3790 mmHg (limits: 0.3751 to 0.3829 mmHg). SvO₂ showed a pre-alignment deviation of 3%, which decreased to 0.8% post-alignment, both within the CLIA threshold of ± 5%, with a Bland-Altman mean difference of 0.7782% (limits: 0.7706 to 0.7858%). Finally, for SaO₂, the pre-alignment deviation was 2.6%, reduced to 0.1% post-alignment, both within the CLIA threshold of ± 5%, with a Bland-Altman mean difference of 0.9614% (limits: 0.9594 to 0.9634%). The Passing-Bablok regression analysis confirmed strong agreement, with slopes close to 1.0100 and intercepts near zero for all parameters. These results validate the QPS as a reliable and non-inferior tool for real-time blood gas monitoring during cardiopulmonary bypass, adhering to CLIA standards and ensuring clinical accuracy.
The findings support the accuracy of the Quantum Perfusion System compared to the BGA standard, demonstrating the system's capability to provide accurate, continuous blood gas monitoring during CPB. However, further studies are necessary to strengthen and confirm these results across broader and more varied clinical scenarios, for these reason as recommended by the manufacturers, the quantum monitoring system should only be used as a trending device.
体外循环(CPB)期间的连续血气监测(CBGM)对于维持患者的最佳预后至关重要,能够对血气参数的关键波动做出快速反应。这项非劣效性研究评估了Spectrum Medical公司的量子灌注系统(Quantum Perfusion System),该系统通过量子工作站(QWS)和量子通气模块(QVM)进行连续在线血气监测,无需使用比色杯,与标准血气分析(BGA)分析仪进行对比,以评估实时临床准确性。
这项回顾性研究纳入了40例患者的样本,使用量子灌注系统进行连续监测,并定期与标准BGA测量结果进行比较。这些患者接受择期CPB手术,具体为冠状动脉旁路移植术(CABG)、二尖瓣置换术(MVR)和主动脉瓣置换术(AVR)。
所有参数的预校准偏差均在临床实验室改进修正案(CLIA)阈值范围内,证实了基线可靠性。对于血红蛋白,预校准偏差为1.9%,校准后降至0.7%,均在CLIA规定的±5%阈值范围内,布兰德-奥特曼平均差异为0.0988 g/dL(范围:0.0963至0.10,12 g/dL)。血细胞比容的预校准偏差为2.1%,校准后降至0.2%,均在CLIA规定的±5%阈值范围内,布兰德-奥特曼平均差异为0.3009%(范围:0.2956至0.3063%)。对于动脉血氧分压(PaO₂),预校准偏差为3.9%,校准后降至0.4%,均在CLIA规定的±10%阈值范围内,布兰德-奥特曼平均差异为4.0490 mmHg(范围:3.9976至4.1004 mmHg)。二氧化碳分压(PCO₂)的预校准偏差为4.2%,校准后降至0.19%,均在CLIA规定的±10%阈值范围内,布兰德-奥特曼平均差异为0.3790 mmHg(范围:0.3751至0.3829 mmHg)。混合静脉血氧饱和度(SvO₂)的预校准偏差为3%,校准后降至0.8%,均在CLIA规定的±5%阈值范围内,布兰德-奥特曼平均差异为0.7782%(范围:0.7706至0.7858%)。最后,对于动脉血氧饱和度(SaO₂),预校准偏差为2.6%,校准后降至0.1%,均在CLIA规定的±5%阈值范围内,布兰德-奥特曼平均差异为0.9614%(范围:0.9594至0.9634%)。帕辛-巴布洛赫回归分析证实了两者之间有很强的一致性,所有参数的斜率接近1.0100,截距接近零。这些结果验证了量子灌注系统作为体外循环期间实时血气监测的可靠且非劣效工具,符合CLIA标准并确保了临床准确性。
研究结果支持量子灌注系统与BGA标准相比的准确性,表明该系统能够在CPB期间提供准确的连续血气监测。然而,有必要进行进一步研究,以在更广泛和更多样化的临床场景中强化和确认这些结果,因此,按照制造商的建议,量子监测系统仅应作为一种趋势监测设备使用。