Sveinsdottir E, Larsen B, Rommer P, Lassen N A
J Nucl Med. 1977 Feb;18(2):168-74.
A computer-based scintillation camera has been designed for both dynamic and static radionuclide studies. The detecting head has 254 independent sodium iodide crystals, each with a photomultiplier and amplifier. In dynamic measurements simultaneous events can be recorded, and 1 million total counts per second can be accommodated with less than 0.5% loss in any one channel. This corresponds to a calculated deadtime of 5 nsec. The multidetector camera is being used for 133Xe dynamic studies of regional cerebral blood flow in man and for 99mTc and 197 Hg static imaging of the brain.
已设计出一种基于计算机的闪烁照相机,用于动态和静态放射性核素研究。检测头有254个独立的碘化钠晶体,每个晶体都配有一个光电倍增管和放大器。在动态测量中,可以记录同时发生的事件,并且每秒可容纳100万个总计数,任何一个通道的损失小于0.5%。这相当于计算得出的死时间为5纳秒。这种多探测器照相机正用于人体局部脑血流量的133Xe动态研究以及大脑的99mTc和197Hg静态成像。