Yuan Junhui, Fang Shaobo, Meng Fan, Wu Yue, Shan Dongqiu, Xu Chunmiao, Zhang Renzhi, Chen Xuejun
Department of Medical Imaging, The Affiliated Cancer Hospital of Zhengzhou University & Henan Cancer Hospital, Zhengzhou, Henan Province, 450008, China.
Department of Radiology, The First Affiliated Hospital of Fujian Medical University, 20 Cha-Zhong Road, Fuzhou, Fujian, 350005, People's Republic of China.
Cancer Imaging. 2025 Apr 8;25(1):53. doi: 10.1186/s40644-025-00867-z.
To investigate the diagnostic value of CE-MATRIX-T1FLAIR and 3D CE-T2FLAIR sequences based on Contrast Enhancement Modulated flip Angle Technique in Refocused Imaging with eXtended echo train (CE-MATRIX) technology for detecting Leptomeningeal Metastasis (LM) using Fluid Attenuated Inversion Recovery (FLAIR) imaging.
This prospective study included 563 hospitalized patients with clinically suspected LM, diagnosed with malignant tumors between January 2022 and October 2023 at Henan Cancer Hospital. Both CE-MATRIX-T1FLAIR and 3D CE-T2FLAIR sequences were used for imaging. Two radiologists independently evaluated image quality, diagnostic confidence, and objective measurements, diagnosing LM as positive or negative, with disagreements resolved by consultation. Subjective and objective scores were compared using the Wilcoxon signed-rank test. The diagnostic performance of the sequences was compared using ROC curve analysis, with cerebrospinal fluid (CSF) cytology as the gold standard. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), accuracy, and area under the curve (AUC) values were calculated and compared using Z-tests.
LM was confirmed in 321 patients. CE-MATRIX-T1FLAIR showed superior subjective scores in image quality and diagnostic confidence (p < 0.001). Though CE-MATRIX-T1FLAIR had a lower SNR (p = 0.013), it demonstrated higher sensitivity, specificity, PPV, NPV, accuracy, and AUC than 3D CE-T2FLAIR (p < 0.001). Both sequences provided effective diagnosis and differentiation of LM.
CE-MATRIX-T1FLAIR offers superior diagnostic performance compared to 3D CE-T2FLAIR for LM, with slightly better subjective ratings despite a lower SNR. Both sequences are effective for diagnosing LM.
探讨基于对比增强调制翻转角技术的CE-MATRIX-T1FLAIR和3D CE-T2FLAIR序列在采用液体衰减反转恢复(FLAIR)成像的重聚焦成像扩展回波链(CE-MATRIX)技术检测软脑膜转移(LM)中的诊断价值。
这项前瞻性研究纳入了563例临床疑似LM的住院患者,这些患者于2022年1月至2023年10月在河南省肿瘤医院被诊断为恶性肿瘤。使用CE-MATRIX-T1FLAIR和3D CE-T2FLAIR序列进行成像。两名放射科医生独立评估图像质量、诊断信心和客观测量结果,将LM诊断为阳性或阴性,如有分歧则通过协商解决。使用Wilcoxon符号秩检验比较主观和客观评分。采用ROC曲线分析比较序列的诊断性能,以脑脊液(CSF)细胞学检查作为金标准。计算并使用Z检验比较敏感性、特异性、阳性预测值(PPV)、阴性预测值(NPV)、准确性和曲线下面积(AUC)值。
321例患者确诊为LM。CE-MATRIX-T1FLAIR在图像质量和诊断信心方面显示出更高的主观评分(p < 0.001)。虽然CE-MATRIX-T1FLAIR的信噪比更低(p = 0.013),但与3D CE-T2FLAIR相比,它具有更高的敏感性、特异性、PPV、NPV、准确性和AUC(p < 0.001)。两种序列均能有效诊断和鉴别LM。
与3D CE-T2FLAIR相比,CE-MATRIX-T1FLAIR对LM具有更高的诊断性能,尽管信噪比更低,但主观评分略好。两种序列对诊断LM均有效。