Abstract

PURPOSE

Pupillary examination is a central part of the neurological assessment. While quantitative pupillometry (QP) improves reliability, the impact of ambient light, particularly on the Neurological Pupil index (NPi), remains unclear. This study aimed to clarify the effects of ambient light on QP parameters in a critical care setting.

METHODS

We performed a prospective crossover study, including 20 adult patients requiring invasive ventilation. Pupillometry was performed during bright condition (BC1), then dark condition (DC), then bright condition again (BC2). In our primary analysis we compared NPi values across conditions (DC1 vs. BC, BC vs. DC2, DC1 vs. DC2). In the secondary analysis, we compared all other QP parameters.

RESULTS

All QP values except constriction velocity and dilation velocity were non-normal. The median NPi was significantly lower in BC compared to dark conditions DC1 in both eyes. In 25% of participants the NPi decreased by 0.6 or more. Conversely, a significant increase in median NPi of both eyes was observed when switching from bright conditions back to dark (BC vs. DC2). No significant difference was found between the two dark condition measurements (DC1 and DC2). The secondary analysis showed that the differences in NPi were driven by differences in most, but not all, QP parameters included in NPi.

CONCLUSIONS

We corroborate previous findings that the level of ambient light affects QP parameters in critically ill patients. This needs to be considered for accurate interpretation of QP parameters. Future studies may explore potential automated light correction methods for wider clinical applicability.

摘要

目的

瞳孔检查是神经学评估的核心部分。虽然定量瞳孔测量法(QP)提高了可靠性,但环境光的影响,尤其是对神经瞳孔指数(NPi)的影响仍不清楚。本研究旨在阐明重症监护环境中环境光对QP参数的影响。

方法

我们进行了一项前瞻性交叉研究,纳入20例需要有创通气的成年患者。在明亮条件下(BC1)进行瞳孔测量,然后在黑暗条件下(DC)进行,之后再次在明亮条件下(BC2)进行。在我们的主要分析中,我们比较了不同条件下的NPi值(DC1与BC、BC与DC2、DC1与DC2)。在次要分析中,我们比较了所有其他QP参数。

结果

除收缩速度和扩张速度外,所有QP值均呈非正态分布。与双眼黑暗条件DC1相比,BC条件下的NPi中位数显著更低。在25%的参与者中,NPi下降了0.6或更多。相反,从明亮条件切换回黑暗条件时(BC与DC2),观察到双眼NPi中位数显著增加。两次黑暗条件测量(DC1和DC2)之间未发现显著差异。次要分析表明,NPi的差异是由NPi中包含的大多数但并非所有QP参数的差异驱动的。

结论

我们证实了先前的研究结果,即环境光水平会影响重症患者的QP参数。在准确解释QP参数时需要考虑这一点。未来的研究可能会探索潜在的自动光校正方法,以实现更广泛的临床应用。