Suppr超能文献

对一些便携式直读式10.2电子伏特和11.8电子伏特光电离探测器以及一台用于高湿度环境中有机蒸汽的火焰离子化气相色谱仪的响应进行评估。

An evaluation of the response of some portable, direct-reading 10.2 eV and 11.8 eV photoionization detectors, and a flame ionization gas chromatograph for organic vapors in high humidity atmospheres.

作者信息

Barsky J B, Que Hee S S, Clark C S

出版信息

Am Ind Hyg Assoc J. 1985 Jan;46(1):9-14. doi: 10.1080/15298668591394293.

Abstract

Three portable, direct-reading instruments were tested to evaluate their ability to detect organic vapors in high humidity atmospheres. The instruments included two photoionization detectors (10.2 eV and 11.8 eV), and an organic vapor analyzer with a flame ionization detector. The organic vapors used were hexanes, 1,1-dichloroethane, chloroform, trichloroethylene, perchloroethylene, benzene, toluene and chlorobenzene. Relatively impervious gas bags were used to contain atmospheres of low (zero) and high (90 percent) relative humidities for comparisons. The response of the organic vapor analyzer was practically independent of humidity. The responses of the photoionization detectors were influenced by high humidity even though the 10.2 eV lamp did not detect water vapor itself. The photoionization detectors were adjudged to be unsuitable for quantitative sampling of humid confined spaces since the 11.8 eV lamp was sensitive to water vapor and the 10.2 eV lamp showed unpredictable response factors in high humidity.

摘要

测试了三种便携式直读仪器,以评估它们在高湿度环境中检测有机蒸汽的能力。这些仪器包括两台光离子化探测器(10.2电子伏特和11.8电子伏特),以及一台配有火焰离子化探测器的有机蒸汽分析仪。所使用的有机蒸汽有己烷、1,1 - 二氯乙烷、氯仿、三氯乙烯、全氯乙烯、苯、甲苯和氯苯。使用相对不透气的气袋来容纳低(零)湿度和高(90%)湿度的大气环境以进行比较。有机蒸汽分析仪的响应实际上与湿度无关。尽管10.2电子伏特的灯本身不能检测水蒸气,但光离子化探测器的响应受到高湿度的影响。由于11.8电子伏特的灯对水蒸气敏感,而10.2电子伏特的灯在高湿度下显示出不可预测的响应因子,因此光离子化探测器被判定不适用于潮湿受限空间的定量采样。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验