Zhu Boyang, Yang Min, Liu Xiaonan, Yuan Bin, Yin Yuqian, Cheng Che, Rehman Faisal Ul
School of Clinical and Basic Medicine, Shandong First Medical University, Shandong, China.
Center for Disease Control and Prevention, Shibei District, Qingdao, Shandong, China.
Clinics (Sao Paulo). 2025 Apr 30;80:100671. doi: 10.1016/j.clinsp.2025.100671. eCollection 2025.
To investigate the practicality, accuracy, and consistency of a computer-based Optokinetic Nystagmus (OKN) analyzer for vision diagnosis in preschoolers.
A comparative study evaluating the performance of a computerized OKN analyzer against the traditional E-word visual acuity chart in assessing visual acuity in preschool children.
A total of 138 children aged 2 to 6 years were enrolled in the study, divided into five age groups: 2 years (29 children), 3 years (27 children), 4 years (28 children), 5 years (28 children), and 6 years (26 children).
Visual acuity was tested using both the nystagmus meter and the E-word visual acuity chart. Statistical analyses, including correlation coefficient calculation and chi-square tests, were conducted to evaluate the testability and consistency of the methods.
Out of 138 participants, 125 (90.6 %) successfully completed the test using the computerized OKN analyzer. There was no significant difference in the measurability across different age groups. The testability of the E-word visual acuity chart was significantly lower at 71.0 %. A strong correlation (R = 0.872) was found between the results from the nystagmus meter and the E-word visual acuity chart, indicating high consistency between the two methods.
The computerized OKN analyzer is a reliable, accurate, and practical tool for assessing visual acuity in preschoolers. It offers a non-invasive alternative to traditional methods, with high testability and strong correlation with standard visual acuity tests, making it a valuable tool for early detection of vision impairments in young children.
研究基于计算机的视动性眼震(OKN)分析仪用于学龄前儿童视力诊断的实用性、准确性和一致性。
一项比较研究,评估计算机化OKN分析仪与传统E字视力表在评估学龄前儿童视力方面的性能。
共有138名2至6岁的儿童参与研究,分为五个年龄组:2岁组(29名儿童)、3岁组(27名儿童)、4岁组(28名儿童)、5岁组(28名儿童)和6岁组(26名儿童)。
使用眼震计和E字视力表测试视力。进行统计分析,包括计算相关系数和卡方检验,以评估这些方法的可测试性和一致性。
在138名参与者中,125名(90.6%)使用计算机化OKN分析仪成功完成测试。不同年龄组之间的可测量性没有显著差异。E字视力表的可测试性显著较低,为71.0%。眼震计和E字视力表的结果之间发现强相关性(R = 0.872),表明两种方法之间具有高度一致性。
计算机化OKN分析仪是评估学龄前儿童视力可靠、准确且实用的工具。它为传统方法提供了一种非侵入性替代方案,具有高可测试性且与标准视力测试有强相关性,使其成为早期发现幼儿视力障碍的有价值工具。