Anetai Yusuke, Tsutsui Yasuhiro, Kinami Shinji, Yokoi Masanori, Tomita Yuji, Koike Yuhei, Takegawa Hideki, Doi Kentaro, Yoshida Ken, Nakamura Satoaki, Yamada Yuji, Nakamura Mitsuhiro
Department of Radiology, Kansai Medical University, 2-5-1 Shin-machi, Hirakata-shi, Osaka, 573-1010, Japan.
Division of Radiation Oncology, Kansai Medical University Hospital, 2-3-1 Shin-machi, Hirakata-shi, Osaka, 573-1191, Japan.
Phys Eng Sci Med. 2025 May 6. doi: 10.1007/s13246-025-01545-x.
Film-specific uniformity variations in packages are known to significantly diminish the effectiveness of the one-scan protocol, a commonly used film dosimetry method. This method universally adopts the reference dose-response with rescaling linearly from the relationship of the known dose and the unexposed state. This study aims to visualize and quantify the variation in unexposed film-specific uniformity in a package to evaluate the suitability of the reference dose response using machine-learning method. Fourteen EBT4 films (#00-#13) were selected from two lot packages. Nine grid-spaced 100 × 100 pixel (72 dpi) patches were obtained from the color images of EBT4 film sheet using a single scanner with landscape (scan A) and portrait (scan B) scan orientations. The reference patch was set at the center of film #00. For this study, multidimensional scaling (MDS) and Lie derivative image analysis (LDIA) were applied to the patch data with respect to the red (R)/green (G)/blue (B) channels. MDS is a suitable method for analyzing non-linear data with similarity, which provides a map of data objects according to a distance metric. LDIA directly detects the deviation vector field between image gradients. The film-specific uniformity was measured at 1/10000 scaled pixel value as a scalar distribution. The image flow field was obtained as a negative gradient of the scalar distribution. Two similarity metrics were defined for comparison with the reference patch: (1) MDSr (the distance parameter in the MDS map from the origin) and (2) Stot (summed S-value in each patch, where S-value represents the vorticity of the deviation vector field obtained via the Lie derivative). MDSr highly correlated with the absolute pixel value difference from the reference patch except for the blue channel in which a favorable package was detected for the reference dose response. Stot quantified the film-uniformity variation from the reference, independent of the dataset, and detected the unfavorable film state as Stot < 0.8 in the blue channel. We visualized and quantified the variation in film-specific uniformity in a lot package using MDS and LDIA, thereby quantitatively determining the unfavorable condition for applying the reference dose-response.
包装中特定胶片的均匀性变化已知会显著降低单次扫描协议的有效性,单次扫描协议是一种常用的胶片剂量测定方法。该方法普遍采用参考剂量响应,并根据已知剂量与未曝光状态的关系进行线性重新缩放。本研究旨在使用机器学习方法可视化和量化包装中未曝光胶片特定均匀性的变化,以评估参考剂量响应的适用性。从两个批次的包装中选取了14张EBT4胶片(#00 - #13)。使用具有横向(扫描A)和纵向(扫描B)扫描方向的单个扫描仪,从EBT4胶片片的彩色图像中获取9个网格间隔为100×100像素(72 dpi)的补丁。参考补丁设置在#00胶片的中心。对于本研究,将多维缩放(MDS)和李导数图像分析(LDIA)应用于关于红色(R)/绿色(G)/蓝色(B)通道的补丁数据。MDS是一种适用于分析具有相似性的非线性数据的方法,它根据距离度量提供数据对象的映射。LDIA直接检测图像梯度之间的偏差向量场。以1/10000缩放像素值作为标量分布来测量特定胶片的均匀性。图像流场作为标量分布的负梯度获得。定义了两个相似性度量以与参考补丁进行比较:(1)MDSr(MDS图中距原点的距离参数)和(2)Stot(每个补丁中的S值总和,其中S值表示通过李导数获得的偏差向量场的涡度)。除了蓝色通道外,MDSr与参考补丁的绝对像素值差异高度相关,在蓝色通道中检测到一个有利于参考剂量响应的包装。Stot量化了相对于参考的胶片均匀性变化,与数据集无关,并在蓝色通道中检测到当Stot < 0.8时胶片状态不利。我们使用MDS和LDIA可视化并量化了批次包装中特定胶片均匀性的变化,从而定量确定了应用参考剂量响应的不利条件。