Average fluorescence intensity is affected by distance and penetration depth with near-infrared fluorescence imaging with indocyanine green in an ex vivo model.

作者信息

Szostek Adrianna K, Grimes Janet A, Wallace Mandy L

出版信息

Am J Vet Res. 2025 May 21;86(8). doi: 10.2460/ajvr.25.01.0001. Print 2025 Aug 1.

Abstract

OBJECTIVE

The objective of this study was to evaluate the effect of indocyanine green (ICG) concentration, distance, angle, and penetration depth on average fluorescence intensity (AFI) when using 2 imaging devices in a canine ex vivo model.

METHODS

3 concentrations of ICG were evaluated (0.006, 0.011, and 0.022 mg/mL) for the effect of distance, penetration depth, angle, and ambient light on AFI. Penetration depth was assessed using serially stacked 1-mm sheets of a gelatin phantom. Each set of conditions was tested using a VITOM 3D and 30° 10-mm laparoscope. Polynomial regression was used to evaluate for significance.

RESULTS

Average fluorescence intensity increased with increasing distance for the 30° 10-mm laparoscope. The VITOM 3D displayed higher AFIs at closer and farther distances. Average fluorescence intensity decreased as penetration depth increased for both devices. Ambient light had no effect on AFI, and AFI varied by angle, but statistical analysis for these variables was limited by insufficient data points.

CONCLUSIONS

Average fluorescence intensity was affected by distance, penetration depth, and angle for each ICG concentration and imaging device. There was no apparent effect on AFI with or without ambient light.

CLINICAL RELEVANCE

The interest in near-infrared fluorescence with ICG is growing in veterinary medicine. Optimization of protocols using these techniques can provide veterinarians with guidelines for the implementation of near-infrared fluorescence with ICG. Further studies are needed to assess the relationship between distance, depth, angle, and ambient light with AFI in vivo.

摘要

目的

本研究的目的是在犬离体模型中使用两种成像设备时,评估吲哚菁绿(ICG)浓度、距离、角度和穿透深度对平均荧光强度(AFI)的影响。

方法

评估了3种ICG浓度(0.006、0.011和0.022 mg/mL)对距离、穿透深度、角度和环境光对AFI的影响。使用明胶模型的连续1毫米厚片评估穿透深度。每组条件均使用VITOM 3D和30°10毫米腹腔镜进行测试。采用多项式回归评估其显著性。

结果

对于30°10毫米腹腔镜,平均荧光强度随距离增加而增加。VITOM 3D在较近和较远的距离显示出更高的AFI。两种设备的平均荧光强度均随穿透深度增加而降低。环境光对AFI没有影响,AFI随角度变化,但这些变量的统计分析因数据点不足而受到限制。

结论

对于每种ICG浓度和成像设备,平均荧光强度受距离、穿透深度和角度的影响。有无环境光对AFI均无明显影响。

临床意义

兽医医学中对ICG近红外荧光的兴趣日益增加。使用这些技术优化方案可为兽医提供实施ICG近红外荧光的指导原则。需要进一步研究以评估体内距离、深度、角度和环境光与AFI之间的关系。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验