Chatborirak Piyapat, Phannarus Krisadi, Kanpittaya Pimsiri, Chantarawaratit Pintu-On
Master of Science Program in Dental Biomaterials Science, Graduate School, Chulalongkorn University, Bangkok, Thailand.
Dental Department, Queen Sirikit National Institute of Child Health, Bangkok, Thailand.
Int Dent J. 2025 May 23;75(4):100834. doi: 10.1016/j.identj.2025.04.014.
Intraoral scanners have recently emerged as a safer and more convenient alternative to traditional impression materials for capturing alveolar morphology in patients with cleft anomalies. This study aimed to assess the accuracy of 3D digital models generated by different intraoral scanners in cleft neonates. Specifically, it focused on evaluating the trueness and precision of these scanners in producing models of unilateral complete cleft lip and palate.
Forty plaster models of unilateral cleft lip and palate from neonates were analyzed. Inter-segment distances, consisting of 37 anterior and 23 posterior parameters, were measured manually with a digital vernier caliper as the reference. The models were scanned using 5 intraoral scanners: 3Shape TRIOS 4, iTero Element 5D, Medit i700, PANDA P2, and Straumann Virtuo Vivo. The differences between tested values (from 3D scans) and reference values were expressed as error per millimeter. Precision was evaluated by performing repeated scans and measurements on 10% of the models. Statistical analyses included 1-sample t-test, Wilcoxon test, and Friedman test, with a significance level of 0.05. Results were further validated through 3D superimposition using Geomagic Control X software.
Observer reliability was excellent, with an ICC value of 1.000. Significant differences in trueness were found for the Medit i700 scanner, with higher errors for posterior inter-segment distances (P = .021) and anterior distances when compared to other scanners (P = .001 to .013). Precision showed excellent agreement across all scanners (ICC = 0.999 to 1.000).
While all scanners met clinical acceptability standards, the Medit i700 demonstrated lower trueness in certain parameters.
This study highlights the importance of scanner selection for accurate digital model generation for cleft neonates.
口腔内扫描仪最近已成为一种比传统印模材料更安全、更便捷的替代方法,用于获取腭裂畸形患者的牙槽形态。本研究旨在评估不同口腔内扫描仪为腭裂新生儿生成的三维数字模型的准确性。具体而言,重点在于评估这些扫描仪在生成单侧完全性唇腭裂模型时的真实性和精确性。
对40个新生儿单侧唇腭裂的石膏模型进行分析。以数字游标卡尺手动测量作为参考,测量由37个前部和23个后部参数组成的节段间距离。使用5种口腔内扫描仪对模型进行扫描:3Shape TRIOS 4、iTero Element 5D、Medit i700、PANDA P2和Straumann Virtuo Vivo。测试值(来自三维扫描)与参考值之间的差异以每毫米误差表示。通过对10%的模型进行重复扫描和测量来评估精确性。统计分析包括单样本t检验、Wilcoxon检验和Friedman检验,显著性水平为0.05。使用Geomagic Control X软件通过三维叠加进一步验证结果。
观察者信度极佳,ICC值为1.000。发现Medit i700扫描仪在真实性方面存在显著差异,与其他扫描仪相比,后部节段间距离的误差更高(P = 0.021),前部距离的误差也更高(P = 0.001至0.013)。所有扫描仪的精确性表现出极佳的一致性(ICC = 0.999至1.000)。
虽然所有扫描仪均符合临床可接受标准,但Medit i700在某些参数上显示出较低的真实性。
本研究强调了为腭裂新生儿选择准确生成数字模型的扫描仪的重要性。