Walburn F J, Sabbah H N, Stein P D
J Biomech Eng. 1985 Aug;107(3):200-5. doi: 10.1115/1.3138544.
The purpose of this study was to measure stresses associated with turbulence (Reynolds stresses), in the region of a 29-mm-dia porcine bioprosthetic valve (Hancock, Model 242). Studies were performed in an in vitro pulse duplicating system with the valve mounted in the aortic position. The Reynolds stresses were calculated from velocities obtained with a two channel laser Doppler anemometer. The largest Reynolds shear stress and normal stress occurred at the highest stroke volume used (80 mL). Averaged over ejection they were 38 dynes/cm2 and 380 dynes/cm2, respectively. The maximal instantaneous Reynolds shear stress was 2500 dynes/cm2 and the maximal instantaneous Reynolds normal stress was 6800 dynes/cm2. Stresses of these magnitudes are in the range reported to damage platelets.
本研究的目的是测量与湍流相关的应力(雷诺应力),该应力存在于直径为29毫米的猪生物瓣膜(汉考克242型)区域内。研究在体外脉动复制系统中进行,瓣膜安装在主动脉位置。雷诺应力由双通道激光多普勒风速仪测得的速度计算得出。最大雷诺剪应力和法向应力出现在使用的最大搏出量(80毫升)时。在射血期平均,它们分别为38达因/平方厘米和380达因/平方厘米。最大瞬时雷诺剪应力为2500达因/平方厘米,最大瞬时雷诺法向应力为6800达因/平方厘米。这些量级的应力处于据报道会损伤血小板的范围内。