Vas Guide: a novel device for looping blood vessels in robot-assisted surgery.

作者信息

Sasaki Yutaro, Fukuta Kyotaro, Yano Tetsuhiro, Nishiyama Mitsuki, Kobayashi Saki, Minato Ryoei, Daizumoto Kei, Tomida Ryotaro, Kusuhara Yoshito, Fukawa Tomoya, Yamaguchi Kunihisa, Yamamoto Yasuyo, Furukawa Junya

机构信息

Department of Urology, Tokushima University Graduate School of Biomedical Sciences, 3-18-15 Kuramoto-cho, Tokushima, 770-8503, Japan.

出版信息

Surg Endosc. 2025 Jul;39(7):4447-4454. doi: 10.1007/s00464-025-11835-w. Epub 2025 Jun 9.

Abstract

BACKGROUND

Looping blood vessels in robot-assisted surgery remains challenging because of limitations in instrument articulation, joint length, and the lack of tactile feedback. To address this critical issue, we developed the Vas Guide, a novel device, for safe blood vessel handling. This study aims to establish the efficacy, safety, and durability of the Vas Guide.

METHODS

The Vas Guide is a flat, curved stainless-steel strip with blunt, rounded ends and a hole near the distal end. After sufficient dissection of the posterior surface of the target blood vessel, the Vas Guide with a vessel loop tied to the distal end was passed around the blood vessel in a single step using a suturing-like motion, thereby enabling secure vessel management. It is reusable and has a unit price of approximately $600. We analyzed 83 patients undergoing robot-assisted partial nephrectomy or nephroureterectomy at Tokushima University Hospital: 39 in the Vas Guide group and 44 in the conventional group. Outcomes including looping time, looping-related complication rates, such as collisions between instruments and bleeding, and additional instrument usage were recorded. The National Aeronautics and Space Administration task load index (NASA-TLX) scores were assessed.

RESULTS

The patients' characteristics were comparable between the two groups. The Vas Guide group demonstrated significantly better outcomes, including faster looping times (31 vs. 90 s, p < 0.001), fewer looping-related complications (0% vs. 30%, p < 0.001), no additional instrument usage (0% vs. 50%, p < 0.001), and better NASA-TLX scores (26.5 vs. 75.7, p < 0.001). The Vas Guide showed no usability issues, with only minor surface scratches and a slight radius change (8.60 mm to 8.69 mm).

CONCLUSION

This reusable device combines high efficacy, safety, and durability and can be implemented at a relatively low cost.

摘要

背景

由于器械关节活动度、关节长度的限制以及缺乏触觉反馈,机器人辅助手术中血管环绕操作仍然具有挑战性。为解决这一关键问题,我们开发了一种新型装置——血管引导器(Vas Guide),用于安全地处理血管。本研究旨在确定血管引导器的有效性、安全性和耐用性。

方法

血管引导器是一条扁平的、弯曲的不锈钢条,两端钝圆,远端附近有一个孔。在充分解剖目标血管的后表面后,将血管环系在远端的血管引导器以类似缝合的动作一步绕过血管,从而实现安全的血管处理。它可重复使用,单价约为600美元。我们分析了在德岛大学医院接受机器人辅助部分肾切除术或肾输尿管切除术的83例患者:血管引导器组39例,传统组44例。记录了包括环绕时间、与环绕相关的并发症发生率(如器械碰撞和出血)以及额外器械使用情况等结果。评估了美国国家航空航天局任务负荷指数(NASA-TLX)得分。

结果

两组患者的特征具有可比性。血管引导器组的结果明显更好,包括更快的环绕时间(31秒对90秒,p < 0.001)、更少的与环绕相关的并发症(0%对30%,p < 0.001)、无额外器械使用(0%对50%,p < 0.001)以及更好的NASA-TLX得分(26.5对75.7,p < 0.001)。血管引导器未出现可用性问题,仅有轻微的表面划痕和轻微的半径变化(从8.60毫米变为8.69毫米)。

结论

这种可重复使用的装置兼具高效、安全和耐用性,且能以相对较低的成本实现。