Solano Antonio, Klein Andrea, Siah Michael C, Gonzalez-Guardiola Gerardo, Chamseddin Khalil, Wagner Aaron, Prakash Vivek, Shih Michael, Baig M Shadman, Timaran Carlos H, Guild Jeffrey, Kirkwood Melissa L
Division of Vascular and Endovascular Surgery, Department of Surgery, University of Texas Southwestern Medical Center, Dallas, TX.
Division of Medical Physics, Department of Radiology, University of Texas Southwestern Medical Center, Dallas, TX.
J Vasc Surg. 2025 Sep;82(3):875-881. doi: 10.1016/j.jvs.2025.05.015. Epub 2025 Jun 7.
Although it is contraindicated for physicians performing fluoroscopically guided interventions (FGIs) to position their hand directly in the beam, it can be unavoidable and results in greater operator exposure and risk. Clinical guidelines recommend against the use of radiation-protective gloves (PGs) during FGIs given the concern for higher radiation accumulation related to increased procedure scatter fraction (PSF). We describe hand radiation dose and procedural scatter during FGIs with regular surgical gloves (RGs) compared to lead-free PGs.
Our primary end point was to evaluate hand radiation dose attenuation of tungsten PG and their effect on operator PSF. In the clinical setting, optically stimulated luminescence detectors were placed bilaterally at the volar and dorsal aspects of the operator's hands at the radioulnar joint and base of the ring finger. Control dots were positioned on the sternum. All operators were trained in applying as low as reasonably achievable principles. In the simulation, fluorography was performed on a 25 × 50 cm, 15-cm thick acrylic phantom for an exposure of 250 mGy reference air kerma. An anthropomorphic hand was placed at 7.5-cm intervals along the length of the table from the point of the fingertips at the center of the x-ray beam, 0 to 30 cm caudally. Radiation dose was calculated based on a calibration coefficient factor to estimate radiation at 80 kVp. The PSF was estimated by normalizing the sternum dose measurement with the dose area product. Wrist and finger radiation doses were normalized by dividing by the sternum dose to control for case length and procedure factors. A paired Wilcoxon test was performed to identify statistically significant differences of normalized PSF, finger, and wrist dose with RG vs PG.
A total of 50 FGIs were performed: 25 with RGs and 25 with PGs. The median procedure reference air kerma was 98 mGy (interquartile range [IQR], 26-137 mGy) with RGs and 63 mGy (IQR, 38-134 mGy) with PGs. The hand radiation dose was lower for both dominant and nondominant hands with the PGs (270 μGy [IQR, 200-520 μGy] vs 590 μGy [IQR, 300-830 μGy], P = .015; and 260 μGy [IQR, 180-240 μGy] vs 660 μGy [IQR, 410-870 μGy]; P < .001) vs RGs. There was no significant increase in PSF with PGs vs RGs (4.46 [IQR, 3.15-5.67] vs 8.21 [IQR, 7.09-12.19]; P = .777). In the simulated setting, hand radiation dose was 58% lower with PG vs RG at all distances (8600 μGy vs 21,000 μGy; P < .001), with no significant differences in PSF for PGs vs RGs.
Operator hands are directly in the x-ray beam more than recognized, leading to an increased lifetime risk. PGs can be expensive and cumbersome, but are effective at decreasing hand radiation exposure without increasing operator radiation dose. They should be considered for use in procedures where hand proximity to the beam is likely.
尽管对于进行透视引导介入操作(FGIs)的医生而言,将手直接置于射线下是禁忌的,但有时却不可避免,这会导致术者接受更多辐射并面临更大风险。鉴于担心与手术散射分数(PSF)增加相关的更高辐射累积,临床指南不建议在FGIs期间使用辐射防护手套(PGs)。我们比较了使用常规手术手套(RGs)与无铅PGs进行FGIs时手部的辐射剂量和手术散射情况。
我们的主要终点是评估钨PG对手部辐射剂量的衰减及其对术者PSF的影响。在临床环境中,将光激励发光探测器双侧放置在术者手部尺桡关节和无名指根部的掌侧和背侧。对照点置于胸骨上。所有术者都接受了尽可能低剂量原则应用的培训。在模拟过程中,在一个25×50 cm、15 cm厚的丙烯酸体模上进行荧光检查,参考空气比释动能为250 mGy。一只仿真手沿着检查台长度从位于X射线束中心的指尖处开始,以7.5 cm的间隔放置,向尾侧0至30 cm。根据校准系数因子计算辐射剂量,以估计80 kVp时的辐射量。通过用剂量面积乘积对胸骨剂量测量值进行归一化来估计PSF。手腕和手指的辐射剂量通过除以胸骨剂量进行归一化,以控制病例长度和手术因素。进行配对Wilcoxon检验,以确定使用RG与PG时归一化PSF、手指和手腕剂量的统计学显著差异。
共进行了50例FGIs:25例使用RGs,25例使用PGs。使用RGs时手术参考空气比释动能的中位数为98 mGy(四分位间距[IQR],26 - 137 mGy),使用PGs时为63 mGy(IQR,38 - 1�4 mGy)。使用PGs时,优势手和非优势手的手部辐射剂量均较低(分别为270 μGy[IQR,200 - 520 μGy]对590 μGy[IQR,300 - 830 μGy],P = 0.015;以及260 μGy[IQR,180 - 240 μGy]对660 μGy[IQR,410 - 870 μGy];P < 0.001),而使用RGs时则较高。与RGs相比,PGs的PSF没有显著增加(分别为4.46[IQR,3.15 - 5.67]对8.21[IQR,7.09 - 12.19];P = 0.777)。在模拟环境中,在所有距离下,使用PG时手部辐射剂量比使用RG时低58%(8600 μGy对21,000 μGy;P < 0.001),PGs与RGs的PSF没有显著差异。
术者手部直接处于X射线下的情况比人们认识到的更为常见,这会导致终生风险增加。PGs可能价格昂贵且使用不便,但能有效减少手部辐射暴露,且不会增加术者的辐射剂量。对于手部可能靠近射线束的操作,应考虑使用PGs。