Sidnell Tim P, Mathias Simone C, Megson David P, Sears Patrick J, Bussemaker Madeleine J
School of Chemistry and Chemical Engineering, University of Surrey, Guildford, UK.
Ecology & Environment Research Centre, Manchester Metropolitan University, Manchester, UK.
Rapid Commun Mass Spectrom. 2025 Sep 30;39(18):e10079. doi: 10.1002/rcm.10079.
Analysis of perfluoroalkyl and polyfluoroalkyl substances (PFAS) is typically achieved using chromatographic techniques, such as high-performance liquid chromatography, combined with mass spectrometry. These techniques can be complex, expensive, time consuming and can lead to inaccurate quantification due to sample container sorption and contamination by polytetrafluoroethylene (PTFE) components.
Here, we demonstrate the novel application of a Direct Probe ionisation Mass Spectrometer (DPiMS) to quantify four aqueous anionic PFAS; perfluorooctanoic acid (PFOA), perfluorooctane sulfonic acid (PFOS), perfluorononanoic acid (PFNA) and 1H,1H,2H,2H-perfluorooctanesulfonic acid (6:2 fluorotelomer surfactant). The DPiMS was operated in both Single Ion Monitoring (SIM) and broad m/z scan range (50-500) modes and the cleaning methodology, ionising voltage and desolvation line heating temperature were optimised.
The derived method quantified both single component and mixtures of PFAS from 0.5 nM to 200 μM (0.2 ppb to 108 ppm) without use of pre-concentration. The advantages of this technique over chromatographic techniques are; the speed of analysis (≈3.5 min per sample, including blanks/cleaning), lack of PTFE components and simplified methodology.
The DPiMS is a useful tool for the rapid screening and estimation of PFAS concentrations with results that are comparable to existing methods. It is anticipated that better sensitivity could be achieved with the use of a triple quadrupole instrument.
全氟烷基和多氟烷基物质(PFAS)的分析通常采用色谱技术,如高效液相色谱,并结合质谱法。这些技术可能复杂、昂贵、耗时,并且由于样品容器吸附和聚四氟乙烯(PTFE)成分的污染,可能导致定量不准确。
在此,我们展示了直接探针电离质谱仪(DPiMS)在定量四种水性阴离子PFAS方面的新应用;全氟辛酸(PFOA)、全氟辛烷磺酸(PFOS)、全氟壬酸(PFNA)和1H,1H,2H,2H-全氟辛烷磺酸(6:2氟调聚物表面活性剂)。DPiMS在单离子监测(SIM)和宽m/z扫描范围(50-500)模式下运行,并对清洗方法、电离电压和脱溶剂线加热温度进行了优化。
所推导的方法无需预浓缩即可定量0.5 nM至200 μM(0.2 ppb至108 ppm)的PFAS单一组分和混合物。该技术相对于色谱技术的优点是;分析速度快(每个样品约3.5分钟,包括空白/清洗)、无PTFE成分且方法简化。
DPiMS是一种用于快速筛选和估算PFAS浓度的有用工具,其结果与现有方法相当。预计使用三重四极杆仪器可实现更高的灵敏度。