使用平台炉塞曼原子吸收光谱法对难分析材料进行金属分析。2. 牛奶中镉和铅的直接测定。

Metal analysis in difficult materials with platform furnace Zeeman-atomic absorption spectroscopy. 2. Direct determination of cadmium and lead in milk.

作者信息

Narres H D, Mohl C, Stoeppler M

出版信息

Z Lebensm Unters Forsch. 1985 Aug;181(2):111-6. doi: 10.1007/BF01042571.

Abstract

A procedure is described for the direct determination of cadmium and lead in whole milk, skim milk, condensed milk, and human milk. Using a Perkin-Elmer 5000 Z instrument with HGA 500 and L'vov platform and by application of oxygen ashing at approx. 600 degrees C, determination limits of approx. 0.02 microgram/l and 0.7 microgram/l for cadmium and lead, respectively, are attainable. Day-to-day precision is 10% for 0.1 microgram/l of cadmium and 2 micrograms/l of lead. Accuracy control at least at higher levels was possible with DPASV after wet digestion. The contents found with this procedure in cows milk are at the lowest limit of very recent literature data, i.e. on average at 0.05 microgram/l for cadmium and 2 microgram/l for lead for samples from nonpolluted regions. The results indicate that milk does not contribute significantly to heavy metal exposure of man.

摘要

本文描述了一种直接测定全脂牛奶、脱脂牛奶、炼乳和人乳中镉和铅含量的方法。使用配备HGA 500和L'vov平台的珀金埃尔默5000 Z型仪器,并在约600摄氏度下采用氧气灰化法,镉和铅的测定限分别可达约0.02微克/升和0.7微克/升。对于0.1微克/升的镉和2微克/升的铅,日常精密度为10%。湿法消解后采用差分脉冲阳极溶出伏安法(DPASV)至少在较高含量水平上进行准确度控制是可行的。用该方法测定的牛奶中镉和铅的含量处于近期文献数据的最低限,即来自无污染地区的样品中,镉平均含量为0.05微克/升,铅平均含量为2微克/升。结果表明,牛奶对人体重金属暴露的贡献不大。