Panhale Vrushali, Kini Raveena, Kothale Shreya
Department of Physiotherapy, Mahatma Gandhi Missions College of Physiotherapy, Plot No: 46, Sector-30, Vashi, Navi Mumbai, 400705, India.
J Hand Microsurg. 2025 Jun 10;17(4):100291. doi: 10.1016/j.jham.2025.100291. eCollection 2025 Jul.
The Jamar dynamometer is the gold standard to assess hand grip strength, but is expensive. The Camry dynamometer is relatively cost-effective. There is a dearth of literature on Camry's validation in healthy Indian adults, which is important to establish as hand grip is known to vary with age, gender and race.
To establish the reliability and validity of the Camry dynamometer in comparison to the Jamar handheld dynamometer among healthy Indian adults in the age group of 20-59 years.
Following basic demographics, occupation and hand dominance, three trials of each dynamometer were performed in standardised positions for the non-dominant and dominant hand of each participant and the best value was recorded. A gap of 10 min was provided between the 2 dynamometers. Data was then recorded and analyzed.
120 participants were recruited, with an equal males and females. The mean isometric hand grip strength for the non-dominant hand was 26.71 ± 9.29 kgf, and for the dominant hand was 27.66 ± 9.12 kgf using the Camry hand-held dynamometer, while it was 26.43 ± 9.07 kgf for the non-dominant and 27.37 ± 9.26 kgf for the dominant side using the Jamar Hand Held dynamometer. Data was further analyzed with age and gender stratification. The Camry hand-held device had excellent reliability (ICC>0.97) and good validity with Pearson's correlation index of 0.97 (p < 0.05) for the dominant hand and 0.99 (p < 0.05) for the nondominant hand, and Bland-Altman's graphics showing more than 90 % of measures within confidence limits.
Camry digital dynamometer is a reliable and valid device to measure isometric handgrip strength in healthy Indian adults compared to the Jamar hydraulic handgrip dynamometer.
Jamar握力计是评估握力的金标准,但价格昂贵。Camry握力计相对性价比高。关于Camry握力计在健康印度成年人中的验证的文献匮乏,鉴于握力已知会因年龄、性别和种族而有所不同,确定这一点很重要。
在20 - 59岁的健康印度成年人中,与Jamar手持式握力计相比,确定Camry握力计的可靠性和有效性。
记录基本人口统计学信息、职业和利手情况后,对每位参与者的非优势手和优势手在标准位置各进行三次每种握力计的测试,并记录最佳值。两种握力计测试之间间隔10分钟。然后记录并分析数据。
招募了120名参与者,男女各半。使用Camry手持式握力计,非优势手的平均等长握力为26.71±9.29千克力,优势手为27.66±9.12千克力;而使用Jamar手持式握力计,非优势手为26.43±9.07千克力,优势手为27.37±9.26千克力。数据按年龄和性别分层进一步分析。Camry手持式设备具有出色的可靠性(组内相关系数>0.97)和良好的有效性,优势手的Pearson相关指数为0.97(P<0.05),非优势手为0.99(P<0.05),Bland - Altman图显示超过90%的测量值在置信限内。
与Jamar液压式握力计相比,Camry数字握力计是测量健康印度成年人等长握力的可靠且有效的设备。