Kuntner Claudia, Friske Joachim, Stessl Andrea, Haas Martin, Breyer Lara Luna, Wanek Thomas, Hacker Marcus, Helbich Thomas, Rausch Ivo
Preclinical Imaging Lab (PIL), Department of Biomedical Imaging and Image-guided Therapy, Medical University of Vienna, Vienna, Austria.
Research Platform Medical Imaging (RPMI), Medical University of Vienna, Vienna, Austria.
EJNMMI Phys. 2025 Jul 1;12(1):63. doi: 10.1186/s40658-025-00771-2.
This study evaluates the performance of a positron emission tomography (PET) insert combined with a 9.4 Tesla magnetic resonance (MR) imaging system using three different radiofrequency (RF) coils. The impact of the RF coils within the PET field of view (FoV) on key PET performance parameters was carefully examined. Additionally, the accuracy of MR-based attenuation correction (AC) on the quantitative PET results was assessed.
The performance of the PET scanner was assessed according to the NEMA NU 4-2008 standards, focusing on sensitivity, spatial resolution, noise-equivalent count (NEC) rate, and image quality metrics (homogeneity, spillover ratios, and recovery coefficients). These evaluations were conducted with three different RF coils within the PET FoV: a mouse-dedicated coil, a rat-dedicated coil (not optimized for PET), and a big-rat-dedicated coil. Quantitative testing of AC accuracy was also performed using homogeneously filled phantoms of varying sizes and activity levels, comparing images reconstructed with and without AC.
The peak absolute sensitivity varied depending on the RF coil used, though spatial resolution was unaffected by the presence of the coils. The NEC curve for the mouse phantom peaked at 441.2 kcps at 29.3 MBq with the mouse-dedicated coil. For the rat phantom, NEC was only tested with the rat-dedicated coils, and the NEC peak reached 203.1 kcps at 27.5 MBq with the big-rat-dedicated coil. Activity concentrations in the NEMA image quality phantom were underestimated when standard MR-AC was applied, likely due to the absence of phantom walls in the generated µ-maps. However, for uniformly filled phantoms with wall thicknesses under 1 mm, standard MR-AC provided accurate quantification.
Overall, the presence of RF coils had only a minor effect on the performance of the PET/MR insert, with the most significant variation observed for the non-PET-optimized coil. Attenuation correction based on generated µ-maps performed well for phantoms with minimal wall thickness.
本研究评估了正电子发射断层扫描(PET)插件与9.4特斯拉磁共振(MR)成像系统结合使用三种不同射频(RF)线圈的性能。仔细研究了PET视野(FoV)内的RF线圈对关键PET性能参数的影响。此外,还评估了基于MR的衰减校正(AC)对定量PET结果的准确性。
根据NEMA NU 4-2008标准评估PET扫描仪的性能,重点关注灵敏度、空间分辨率、噪声等效计数(NEC)率和图像质量指标(均匀性、溢出率和恢复系数)。这些评估是在PET FoV内使用三种不同的RF线圈进行的:小鼠专用线圈、大鼠专用线圈(未针对PET进行优化)和大大鼠专用线圈。还使用不同尺寸和活性水平的均匀填充体模对AC准确性进行了定量测试,比较了有AC和无AC重建的图像。
峰值绝对灵敏度因使用的RF线圈而异,尽管空间分辨率不受线圈存在的影响。小鼠体模的NEC曲线在使用小鼠专用线圈时,在29.3 MBq时达到441.2 kcps的峰值。对于大鼠体模,仅使用大鼠专用线圈测试了NEC,使用大大鼠专用线圈时,NEC峰值在27.5 MBq时达到203.1 kcps。当应用标准MR-AC时,NEMA图像质量体模中的活性浓度被低估,这可能是由于生成的µ图中没有体模壁。然而,对于壁厚小于1毫米的均匀填充体模,标准MR-AC提供了准确的定量。
总体而言,RF线圈的存在对PET/MR插件的性能影响较小,对于未针对PET优化的线圈观察到的变化最为显著。基于生成的µ图的衰减校正对于壁厚最小的体模表现良好。