因连接处过度弯曲导致带球形接头浮动分离环的线圈分离术后粘连:一例报告及台架实验

Sticking after detachment procedure of coil with floating detachment link with ball joint due to excessively bent junction: A case report and a bench-top experiment.

作者信息

Akaike Natsuki, Ikeda Hiroyuki, Tanimura Mai, Uezato Minami, Kinosada Masanori, Kurosaki Yoshitaka, Chin Masaki

机构信息

Department of Neurosurgery, Kurashiki Central Hospital, Okayama, Japan.

出版信息

Surg Neurol Int. 2025 Jun 13;16:244. doi: 10.25259/SNI_366_2025. eCollection 2025.

Abstract

BACKGROUND

We present a case of coil sticking, likely caused by an excessively bent junction of the coil and the delivery pusher, with a bench-top experiment to determine the cause of coil sticking in this case.

CASE DESCRIPTION

A coil embolization procedure was performed for an internal carotid artery aneurysm. During the framing with the Axium Prime Frame coil, the tip of the microcatheter was advanced to the center of the coil mass within the aneurysm to insert the coil. After detachment, the coil became stuck. Pulling the delivery pusher allowed the coil to detach from the delivery pusher within the microcatheter. After flushing the microcatheter with heparinized saline, the proximal end of the coil fortuitously settled within the aneurysm and the procedure was concluded without complications. A bench-top experiment suggested that an excessively bent junction of the coil and the delivery pusher causes the ball joint of the floating detachment link to hook into the circular opening.

CONCLUSION

When using coils with a floating detachment link that has a ball joint, forcibly advancing the microcatheter while the tip of the microcatheter is aligned with the junction may cause excessive bending of the junction, resulting in coil sticking.

摘要

背景

我们报告一例弹簧圈粘连病例,可能是由于弹簧圈与输送推送器的连接处过度弯曲所致,并通过一个台式实验来确定该病例中弹簧圈粘连的原因。

病例描述

对一例颈内动脉瘤进行了弹簧圈栓塞手术。在使用Axium Prime Frame弹簧圈成篮过程中,将微导管尖端推进至动脉瘤内弹簧圈团块的中心以插入弹簧圈。分离后,弹簧圈发生粘连。拉动输送推送器使弹簧圈在微导管内与输送推送器分离。用肝素盐水冲洗微导管后,弹簧圈近端偶然落入动脉瘤内,手术顺利完成,无并发症。一个台式实验表明,弹簧圈与输送推送器的连接处过度弯曲会导致浮动分离链接的球形接头钩入圆形开口。

结论

当使用带有球形接头的浮动分离链接的弹簧圈时,在微导管尖端与连接处对齐时强行推进微导管可能会导致连接处过度弯曲,从而导致弹簧圈粘连。