Multi-centric clinical implementation of the remote and automated quality control programme for digital imaging in Malaysia: challenges and pitfalls.

作者信息

Ravichandran Vinoshah S, Hamzah Nur Ammi, Tan Li Kuo, Tsapaki Virginia, Bjelac Olivera Ciraj, Mohd Noor Noramaliza, Mohamad Mokhtar Nur Syahiirah, Abd Rashid Nur Hasyimah, Saiful Fazad Adiela, Zakaria Nur Hafizah, Mohd Mystafa Siti Norsyafiqah, Wan Ghazali Wan Nur Ain, Mohd Yusoff Nur Shahidatul Akma, Khalid Norafatin, Yeong Chai Hong, Abdul Karim Muhammad Khalis, Wong Jeannie Hsiu Ding

机构信息

Department of Biomedical Imaging, Faculty of Medicine, Universiti Malaya, 50603, Kuala Lumpur, Malaysia.

Medical Radiation Surveillance Unit, Public Health Department, Melaka State Health Department, Ayer Keroh, Malaysia.

出版信息

Phys Eng Sci Med. 2025 Jul 17. doi: 10.1007/s13246-025-01590-6.

Abstract

The International Atomic Energy Agency (IAEA) has developed a methodology for a remote and automated quality control (QC) programme for digital radiography (DR) units. The purpose of this paper is to report the results of the implementation of the methodology in four hospitals in Malaysia. The IAEA methodology provides multiple image quality metrics by using dedicated software and standard, easily available materials to construct phantoms at a reasonably low cost. Nine QC phantoms were constructed and distributed across these institutions, with data collected daily or weekly analysed using the Python implementation of the Automated Tool for Image Analysis software. Image quality metrics, including signal-difference-to-noise ratio (SDNR), signal-noise ratio (SNR), modulation transfer function (MTF) and detectability index (d') were assessed on 11 digital radiography units from four different manufacturers. The use of diverse imaging protocols resulted in statistically significant differences in all the image quality metrics across the different units. For the processed image protocols, the median SDNR values ranged (12.2-17.5) and (9.1-17.9), respectively and were less affected by the protocol variations compared to SNR values. The d' 0.3 mm ranged (4.8-7.1) and (3.4-6.2), while the d' 4 mm variation ranged (73-115) and (83-130), respectively. The MTF values were strongly correlated between the horizontal and vertical MTFs, as well as between the MTF levels at 10%, 20% and 50%. Across different DR units, there were significant differences in the image quality metrics, mainly due to the different acquisition protocols employed. Clinical protocols have inherent image post-processing that can significantly alter the image quality metrics values compared to the raw image. The IAEA methodology is a useful tool to track the performance of DR units over time. Recommendations for the wider implementation of this methodology would include standardising the acquisition protocol by means of setting a specific QC protocol template to ensure consistency in the image acquisition.

摘要

国际原子能机构(IAEA)已开发出一种用于数字射线摄影(DR)设备的远程自动化质量控制(QC)程序的方法。本文旨在报告该方法在马来西亚四家医院的实施结果。国际原子能机构的方法通过使用专用软件和标准的、易于获取的材料以合理的低成本构建体模,提供多种图像质量指标。构建了九个质量控制体模并分发给这些机构,使用图像分析软件自动化工具的Python实现对每日或每周收集的数据进行分析。在来自四个不同制造商的11台数字射线摄影设备上评估了图像质量指标,包括信号差噪比(SDNR)、信噪比(SNR)、调制传递函数(MTF)和可探测性指数(d')。不同成像协议的使用导致不同设备的所有图像质量指标存在统计学上的显著差异。对于处理后的图像协议,SDNR中位数分别在(12.2 - 17.5)和(9.1 - 17.9)范围内,与SNR值相比,受协议变化的影响较小。d' 0.3毫米范围分别为(4.8 - 7.1)和(3.4 - 6.2),而d' 4毫米变化范围分别为(73 - 115)和(83 - 130)。MTF值在水平和垂直MTF之间以及10%、20%和50%的MTF水平之间具有很强的相关性。在不同的DR设备之间,图像质量指标存在显著差异,主要是由于采用了不同的采集协议。与原始图像相比,临床协议具有固有的图像后处理功能,可显著改变图像质量指标值。国际原子能机构的方法是一种随着时间推移跟踪DR设备性能的有用工具。更广泛实施该方法的建议包括通过设置特定的质量控制协议模板来标准化采集协议,以确保图像采集的一致性。