Mashego Portia Tshidi, Hooijberg Emma H
Department of Companion Animal Clinical Studies, Faculty of Veterinary Science, University of Pretoria, Pretoria, Gauteng, South Africa.
Vet Clin Pathol. 2025 Sep;54(3):239-250. doi: 10.1111/vcp.70037. Epub 2025 Jul 18.
An immunoturbidometric assay for symmetric dimethylarginine (SDMA) measurement on automated chemistry analyzers has recently become available.
To perform analytical validation of the EUROLyser SDMA assay in dogs and cats.
Method validation experiments were performed using stored canine and feline serum. Quality control validation was performed according to Westgard. Performance goals were derived from SDMA biological variation data for both species.
Imprecision ranged from 3.7%-7.8% (dogs) to 6.0%-11.8% (cats) with a dispersion of ±35% for dogs and ±44% for cats. The assay showed linearity (up to 85 μg/dL [dogs], 75 μg/dL [cats]); the preliminary LoQ was 9.5 and 6.9 μg/dL, respectively. Recovery was 19.7% and 6.5%, respectively. Severe hemolysis resulted in a significant bias in both species. The EUROLyser method showed a significant negative proportional and constant bias in dogs and a significant positive proportional and negative constant bias in cats, compared to the comparative method. In dogs, the mean bias (-19.5%) and the bias at clinical decision limits exceeded the desirable bias; in cats, the mean bias (-4.0%) and the bias at clinical decision limits was < 8%. Because of high imprecision, the TE that could be controlled for was 35%, with a 1-2.5 s rule using patient pools as quality control material.
The novel SDMA assay showed acceptable analytical performance, but high dispersion has consequences for the interpretation of results at reference limits and serial measurements. Method-specific reference intervals and decision limits should be generated for both species, particularly for dogs, given the significant bias vs. the comparative method.
一种用于在自动化学分析仪上测量对称二甲基精氨酸(SDMA)的免疫比浊法最近已可用。
对犬猫的EUROLyser SDMA检测方法进行分析验证。
使用储存的犬猫血清进行方法验证实验。根据韦斯特加德方法进行质量控制验证。性能目标源自两个物种的SDMA生物学变异数据。
不精密度范围为3.7% - 7.8%(犬)至6.0% - 11.8%(猫),犬的离散度为±35%,猫为±44%。该检测方法显示出线性(犬可达85μg/dL,猫可达75μg/dL);初步定量下限分别为9.5和6.9μg/dL。回收率分别为19.7%和6.5%。严重溶血在两个物种中均导致显著偏差。与比较方法相比,EUROLyser方法在犬中显示出显著的负比例偏差和恒定偏差,在猫中显示出显著的正比例偏差和负恒定偏差。在犬中,平均偏差(-19.5%)和临床决策限处的偏差超过了理想偏差;在猫中,平均偏差(-4.0%)和临床决策限处的偏差<8%。由于不精密度高,使用患者样本池作为质量控制材料时,可控制的总误差为35%,采用1 - 2.5s规则。
新型SDMA检测方法显示出可接受的分析性能,但高离散度对参考限和连续测量结果的解释有影响。鉴于与比较方法存在显著偏差,应为两个物种,特别是犬,制定特定方法的参考区间和决策限。