Masson Ravi, Sineri Charles, Cheung Edwin, Reilly John, Landau David, Chen On
Division of Cardiology, Department of Medicine, Stony Brook University Hospital, Stony Brook, NY.
Division of Vascular and Endovascular Surgery, Department of Surgery, Stony Brook University Hospital, Stony Brook, NY.
J Vasc Surg Cases Innov Tech. 2025 Jun 4;11(5):101868. doi: 10.1016/j.jvscit.2025.101868. eCollection 2025 Oct.
Permanent vascular closure systems are commonly used for achieving hemostasis after transfemoral percutaneous procedures. However, there exists conflicting data regarding the safety of re-puncturing these devices during subsequent procedures. Because percutaneous procedures have become more prevalent, so have the use of vascular closure devices, emphasizing the need to understand their limitations. The following case report describes a situation where a micropuncture catheter fractured within a closure device during inadvertent repeat device access and the challenges encountered with its retrieval.