Kuchnicki T C, Campbell N E
Anal Biochem. 1985 Aug 15;149(1):111-6. doi: 10.1016/0003-2697(85)90482-8.
Hydrogen-conditioned Clark electrodes exposed to successive acetylene samples displayed a decreasing curvilinear response. Slight electrode responses are also effected by saturated ethylene solutions. Thus, amperometric measurements of hydrogen or carbon monoxide in the aqueous environment by these electrodes are perturbated by the presence of acetylene and to a much lesser extent by ethylene. Methane-saturated solutions induced no electrode response.
暴露于连续乙炔样品的氢调节克拉克电极呈现出递减的曲线响应。饱和乙烯溶液也会对电极产生轻微响应。因此,这些电极在水性环境中对氢气或一氧化碳的安培测量会受到乙炔的干扰,而受到乙烯的干扰程度要小得多。甲烷饱和溶液未引起电极响应。