Chin Taeyang, Padalko Elizaveta
Department of Medical Microbiology, Ghent University Hospital, Ghent, Belgium.
Acta Clin Belg. 2025 Aug;80(4):83-90. doi: 10.1080/17843286.2025.2543336. Epub 2025 Aug 1.
To evaluate the analytical performance of the automated AIX1000 RPR assay compared to the manual Macro-Vue RPR Card Test for syphilis monitoring in a high-prevalence, reverse screening setting.
Serum samples were tested using both AIX1000 and manual RPR in retrospective ( = 75) and prospective ( = 279) cohorts. Qualitative and quantitative concordance were assessed, along with AIX1000's accuracy, precision, specificity, carry-over, and freeze-thaw stability.
In the retrospective cohort, overall qualitative agreement was 90.7% (κ = 0.80), with 77.3% of samples showing titers within one dilution. In the prospective cohort, qualitative agreement was 86.0% (κ = 0.72), with 90.7% of samples within one dilution. AIX1000 results were often 1-2 dilutions lower than manual RPR, especially in high-titer samples (≥1:16), where titer concordance (±1 dilution step) dropped to 38.7% (prospective) and 35.3% (retrospective). Accuracy analysis using external quality controls revealed a consistent one dilution-step bias for both assays: manual RPR read higher and AIX1000 lower, explaining the observed discrepancies. Analytical performance met predefined criteria: within and between runs were consistent, specificity showed no cross-reactivity, and no carry-over contamination was observed. Freeze-thaw testing had minimal effect on results.
The AIX1000 showed good concordance with manual RPR, particularly at lower titers, with reliable analytical performance and operational advantages. However, systematic titer underestimation compared to manual RPR, particularly in high-titer samples, may affect clinical interpretation if tests are transitioned. Clinicians should be informed of this discrepancy when transitioning platforms.
在高流行率的反向筛查环境中,评估自动化AIX1000快速血浆反应素环状卡片试验(RPR)与手动Macro-Vue RPR卡片试验在梅毒监测方面的分析性能。
在回顾性队列(n = 75)和前瞻性队列(n = 279)中,使用AIX1000和手动RPR对血清样本进行检测。评估定性和定量一致性,以及AIX1000的准确性、精密度、特异性、携带污染和冻融稳定性。
在回顾性队列中,总体定性一致性为90.7%(κ = 0.80),77.3%的样本滴度在一个稀释度范围内。在前瞻性队列中,定性一致性为86.0%(κ = 0.72),90.7%的样本在一个稀释度范围内。AIX1000的结果通常比手动RPR低1 - 2个稀释度,特别是在高滴度样本(≥1:16)中,滴度一致性(±1个稀释度步长)降至38.7%(前瞻性)和35.3%(回顾性)。使用外部质量控制进行的准确性分析显示,两种检测方法均存在一致的一个稀释度步长偏差:手动RPR读数较高,AIX1000读数较低,这解释了观察到的差异。分析性能符合预定义标准:批内和批间一致,特异性无交叉反应,未观察到携带污染。冻融试验对结果影响最小。
AIX1000与手动RPR显示出良好的一致性,尤其是在较低滴度时,具有可靠的分析性能和操作优势。然而,与手动RPR相比,系统性滴度低估,特别是在高滴度样本中,如果转换检测方法,可能会影响临床解读。在转换平台时,应告知临床医生这种差异。