Biocompatibility of three orthodontic silver solder materials: in vitro cytotoxicity testing.

作者信息

Battula Sandeep, Aileni Kaladhar Reddy, Rachala Madhukar Reddy, Dasari Arun Kumar, Ganji Kiran Kumar

机构信息

Department of Orthodontics, SVS Institute of Dental Sciences, Mahabubnagar, Telangana, India.

Department of Preventive Dentistry, College of Dentistry, Jouf University, Sakaka, Al Jouf, Saudi Arabia.

出版信息

J Orofac Orthop. 2025 Aug 15. doi: 10.1007/s00056-025-00608-8.

Abstract

OBJECTIVE

To comparatively evaluate the in vitro cytotoxicity of three commercially available silver solder materials in different forms-traditional silver solder with flux, silver solder incorporated with flux, and silver solder paste with flux-on Balb/c 3T3 fibroblast cell lines.

MATERIALS AND METHODS

Ninety-six samples were categorized into four groups: group 1 (only cell culture medium), group 2 (silver solder + flux), group 3 (universal silver solder), and group 4 (silver solder paste with flux). Each group had 24 samples assessed at three intervals: 24 h (T1), 7 days (T2), and 30 days (T3). Cytotoxicity was evaluated using the neutral red uptake assay, measuring optical density at 540 nm. Data were analyzed using one-way analysis of variance (ANOVA), Tukey's post hoc test, and paired t‑tests (p < 0.05).

RESULTS

All test groups exhibited increased cytotoxicity over time, with the highest toxicity observed at 24 h. Group 4 (solder paste with flux) showed the greatest cytotoxicity (96.75% nonviable cells), followed by group 2 (76.13%) and group 3 (71.83%). The control group showed minimal change (16.4%). ANOVA revealed significant differences between groups at all-time points (p < 0.001). Post hoc analysis confirmed that the solder paste group was significantly more cytotoxic than the others. Time-dependent reductions in cell viability were observed, particularly between T2 and T3.

CONCLUSION

All tested silver solder materials demonstrated cytotoxic effects on Balb/c 3T3 fibroblasts. The silver solder and flux displayed the lowest cytotoxicity, followed by universal silver solder, and silver solder paste with flux demonstrating the highest cytotoxicity. These findings highlight the need for careful material selection in orthodontic appliance fabrication to minimize adverse biological effects.

摘要

目的

比较评估三种市售不同形态的银焊料材料——含助焊剂的传统银焊料、含助焊剂的银焊料以及含助焊剂的银焊膏——对Balb/c 3T3成纤维细胞系的体外细胞毒性。

材料与方法

96个样本分为四组:第1组(仅细胞培养基)、第2组(银焊料+助焊剂)、第3组(通用银焊料)和第4组(含助焊剂的银焊膏)。每组有24个样本,在三个时间点进行评估:24小时(T1)、7天(T2)和30天(T3)。使用中性红摄取试验评估细胞毒性,测量540nm处的光密度。数据采用单因素方差分析(ANOVA)、Tukey事后检验和配对t检验进行分析(p<0.05)。

结果

所有测试组的细胞毒性均随时间增加,在24小时时观察到最高毒性。第4组(含助焊剂的焊膏)显示出最大的细胞毒性(96.75%的非存活细胞),其次是第2组(76.13%)和第3组(71.83%)。对照组变化最小(16.4%)。方差分析显示所有时间点组间存在显著差异(p<0.001)。事后分析证实,焊膏组的细胞毒性明显高于其他组。观察到细胞活力随时间下降,尤其是在T2和T3之间。

结论

所有测试的银焊料材料均对Balb/c 3T3成纤维细胞表现出细胞毒性作用。银焊料和助焊剂的细胞毒性最低,其次是通用银焊料,含助焊剂的银焊膏细胞毒性最高。这些发现凸显了在正畸矫治器制作中仔细选择材料以尽量减少不良生物学效应的必要性。