Ülfer Gözde
Istanbul Medipol University, Faculty of Medicine, Department of Biochemistry, Istanbul, Türkiye.
J Med Biochem. 2025 Jun 13;44(3):479-485. doi: 10.5937/jomb0-56100.
This study aimed to compare the prothrombin time (PT), international normalised ratio (INR) and activated partial thromboplastin time (aPTT) values obtained using the photo-optical method and to assess these values according to the reference method, which was the mechanical method.
Plasma samples from 340 patients, submitted to our hospital's biochemistry laboratory for PT, INR, and aPTT analyses, were assayed using the mechanical coagulometric measurement method in a Stago Compact Max3 automated coagulation analyser, which served as the reference device. The same samples were also analysed using the Sunbio UP5500 automated analyser with a simultaneous optical method. There were 30 turbid samples analysed in both devices without exclusion from the study. Correlation coefficient analysis was carried out using SPSS to assess intervariable correlations. Passing-Bablok regression analysis was performed in R software version 3.6.0 to compare PT, INR, and aPTT values between the two devices. Bland-Altman plots were used to analyse the agreement.
A good level of statistically significant agreement was found between the PT and INR values measured by the Stago Compact Max3 and Sunbio UP 5500 devices (Interclass Coefficient Correlation (ICC): 0.627, p=0.001; p<0.01 and ICC: 0.653, p=0.001; p<0.01, respectively). Additionally, there was an excellent level of statistically significant agreement for the aPTT values (ICC: 0.902, p=0.001, p<0.01). The Bland-Altman analysis revealed the mean 95% limits of agreement values as 2.46 (lower limit: -2.44, upper limit: 7.37) for PT, 0.07 (lower limit: -0.32, upper limit: 0.46) for INR, and 2.45 (lower limit: -1.67, upper limit: 6.58) for aPTT. The Passing-Bablok regression results indicated a systematic difference for PT measurement but no proportional difference. No systematic or proportional differences were found for the measured INR and aPTT values between the Stago Compact Max3 and Sunbio UP 5500 devices. The intra-assay and interassay coefficient of variation (CV) values from level 1 and 2 controls of the optical method were below 5%.
The results from the optical method were consistent and reliable compared to the mechanical method. PT and INR results showed statistically good agreement, while aPTT results demonstrated excellent agreement. Larger multicenter studies are needed to evaluate turbid samples.
本研究旨在比较采用光学方法获得的凝血酶原时间(PT)、国际标准化比值(INR)和活化部分凝血活酶时间(aPTT)值,并根据作为参考方法的机械方法来评估这些值。
将340例患者的血浆样本送至我院生化实验室进行PT、INR和aPTT分析,使用Stago Compact Max3自动凝血分析仪中的机械凝固法测量方法进行检测,该分析仪作为参考设备。同样的样本也使用具有同步光学方法的Sunbio UP5500自动分析仪进行分析。在两台设备中对30份浑浊样本进行了分析,且未将其排除在研究之外。使用SPSS进行相关系数分析以评估变量间的相关性。在R软件版本3.6.0中进行Passing-Bablok回归分析,以比较两台设备之间的PT、INR和aPTT值。使用Bland-Altman图分析一致性。
Stago Compact Max3和Sunbio UP 5500设备测量的PT和INR值之间发现了具有统计学意义的良好一致性水平(组内相关系数(ICC):0.627,p = 0.001;p < 0.01和ICC:0.653,p = 0.001;p < 0.01,分别)。此外,aPTT值具有统计学意义的出色一致性水平(ICC:0.902,p = 0.001,p < 0.01)。Bland-Altman分析显示,PT的平均95%一致性界限值为2.46(下限:-2.44,上限:7.37),INR为0.07(下限:-0.32,上限:0.46),aPTT为-2.45(下限:-1.67,上限:6.58)。Passing-Bablok回归结果表明PT测量存在系统差异,但无比例差异。在Stago Compact Max3和Sunbio UP 5500设备之间测量的INR和aPTT值未发现系统或比例差异。光学方法1级和2级对照的批内和批间变异系数(CV)值均低于5%。
与机械方法相比,光学方法的结果一致且可靠。PT和INR结果显示出统计学上的良好一致性,而aPTT结果显示出出色的一致性。需要更大规模的多中心研究来评估浑浊样本。