Dorn-Beineke Alexandra, Dempfle Carl-Erik, Bertsch Thomas, Wisser Herrmann
Institute for Clinical Chemistry, Faculty for Clinical Medicine Mannheim, University of Heidelberg, Theodor-Kutzer-Ufer 1-3, Mannheim 68167, Germany.
Thromb Res. 2005;116(2):171-9. doi: 10.1016/j.thromres.2004.12.001. Epub 2005 Jan 28.
Centralization of laboratory diagnostics and an increasing number of urgent requests and nonstandard samples raise the need for short turn-around times and high-throughput analyzers for coagulation tests. The aim of the present study was to evaluate the analytical and technical performance of the Sysmex CA-7000 coagulation analyzer under routine laboratory conditions.
We evaluated the Sysmex CA-7000 in comparison to the Sysmex CA-6000 analyzer for PT, INR, aPTT, Clauss, and derived fibrinogen. We also compared antithrombin (AT) measured on the Sysmex CA-7000 and Dimension RxL. Imprecision studies were performed, and throughput, online, and STAT functions and the handling of routine samples were evaluated.
The Sysmex CA-7000 showed very low intra-assay and interassay variability for all parameters. The method comparison study showed good comparability to the Sysmex CA-6000 and Dimension RxL. The throughput of the Sysmex CA-7000 was 2.5--3 times faster than that of Sysmex CA-6000. No interference was seen for total bilirubin up to 710 micromol/L. Triglyceride levels >11.4 mmol/L resulted in invalid PT measurements and levels >3.42 mmol/L gave invalid results for Clauss fibrinogen on the CA-7000 analyzer.
This is the first published evaluation of the analytical and technical performance of the coagulation analyzer Sysmex CA-7000. We conclude that the CA-7000 is well suited for coagulation laboratories with high sample throughput and a high number of nonstandard samples.
实验室诊断的集中化以及紧急检测请求和非标准样本数量的增加,使得凝血检测需要更短的周转时间和高通量分析仪。本研究的目的是评估Sysmex CA - 7000凝血分析仪在常规实验室条件下的分析和技术性能。
我们将Sysmex CA - 7000与Sysmex CA - 6000分析仪进行比较,检测凝血酶原时间(PT)、国际标准化比值(INR)、活化部分凝血活酶时间(aPTT)、Clauss法纤维蛋白原及衍生纤维蛋白原。我们还比较了在Sysmex CA - 7000和Dimension RxL上检测的抗凝血酶(AT)。进行了精密度研究,并评估了通量、在线和即时检测(STAT)功能以及常规样本的处理情况。
Sysmex CA - 7000在所有参数上均显示出非常低的批内和批间变异。方法比较研究表明,其与Sysmex CA - 6000和Dimension RxL具有良好的可比性。Sysmex CA - 7000的通量比Sysmex CA - 6000快2.5至3倍。总胆红素高达710微摩尔/升时未观察到干扰。甘油三酯水平>11.4毫摩尔/升会导致CA - 7000分析仪上PT测量结果无效,而水平>3.42毫摩尔/升会使Clauss法纤维蛋白原检测结果无效。
这是首次发表的关于凝血分析仪Sysmex CA - 7000分析和技术性能的评估。我们得出结论,CA - 7000非常适合样本通量高且非标准样本数量多的凝血实验室。