Dusi Francesca, Busato Fabio, Testolin Alessandro, Fiorentin Davide, Tessari Francesca, Nicoletto Mattia, Fusella Marco
Abano Terme Hospital, Department of Radiation Oncology, Abano Terme, Padua, Italy.
Abano Terme Hospital, Department of Radiation Oncology, Abano Terme, Padua, Italy.
Phys Med. 2025 Sep;137:105091. doi: 10.1016/j.ejmp.2025.105091. Epub 2025 Aug 23.
BACKGROUND/INTRODUCTION: This study evaluates the dosimetric performance of HyperSight CBCT (HS-CBCT) images for CBCT-only adaptive radiotherapy workflows.
We evaluated Hounsfield unit (HU) accuracy and dose calculation performance using both Feldkamp-Davis-Kress (FDK) and Acuros reconstructions in phantom and clinical datasets (brain, breast, pelvis). HU deviations were reported relative to planning CT. Longitudinal HU stability was assessed over 20 CBCT sessions in a prostate cohort. Gamma passing rates (2 %/2mm) were used for dosimetric evaluation.
Acuros-reconstructed CBCT images demonstrated HU agreement within 28 HU of the planning CT, compared to deviations up to 284 HU for FDK (the relative coefficient of variation -CV- of the difference from the pCT reference value were 2 % and 20 % for Acuros and FDK respectively). Dose recalculations on Acuros images resulted in higher gamma passing rates, particularly in pelvic cases (97.8 % vs 96.6 % with FDK, p < 0.05). HU stability analysis over 20 CBCT sessions in prostate patients confirmed reproducibility, with standard deviations of 5.0 HU for bone and 4.2 HU for bladder.
This study provides the first comprehensive, multi-site and longitudinal clinical validation of HyperSight CBCT with Acuros reconstruction for ART workflows. Our results demonstrate high HU fidelity and robust dose calculation performance, supporting the clinical feasibility of CBCT-only workflows across different anatomical regions and throughout the course of treatment.
背景/引言:本研究评估了用于仅CBCT自适应放疗流程的HyperSight CBCT(HS-CBCT)图像的剂量学性能。
我们在体模和临床数据集(脑、乳腺、骨盆)中使用Feldkamp-Davis-Kress(FDK)和Acuros重建方法评估了Hounsfield单位(HU)准确性和剂量计算性能。报告了相对于计划CT的HU偏差。在一个前列腺队列中,对20次CBCT扫描的HU纵向稳定性进行了评估。使用γ通过率(2%/2mm)进行剂量学评估。
与FDK高达284 HU的偏差相比,Acuros重建的CBCT图像显示与计划CT的HU一致性在28 HU以内(Acuros和FDK与pCT参考值差异的相对变异系数-CV-分别为2%和20%)。对Acuros图像进行剂量重新计算可得到更高的γ通过率,尤其是在盆腔病例中(97.8%对FDK的96.6%,p<0.05)。对前列腺患者进行的20次CBCT扫描的HU稳定性分析证实了其可重复性,骨的标准差为5.0 HU,膀胱的标准差为4.2 HU。
本研究首次对用于自适应放疗流程的采用Acuros重建的HyperSight CBCT进行了全面、多部位和纵向的临床验证。我们的结果显示了高HU保真度和稳健的剂量计算性能,支持了在不同解剖区域以及整个治疗过程中仅使用CBCT流程的临床可行性。