Gressner A M, Henn K H
J Clin Chem Clin Biochem. 1985 Nov;23(11):781-5. doi: 10.1515/cclm.1985.23.11.781.
The enzymatic determination of sialic acid, commercially introduced recently as an end point method, has been considerably improved by the development of initial rate and peak rate modifications of the kinetic method, respectively, and adapted to a centrifugal analyser (Cobas Bio). The assay is completed within 3 min, takes 3 microliter sample volume and is reduced to a total reaction volume of 153 microliter. The costs per test thereby are lowered to about 10% of the original procedure. Within-run and day-to-day imprecisions (means = 0.81 g/l) are 2% and 3.9%, respectively; inaccuracy is between +1% and -2%. The detection limit is 0.012 g/l, i.e. about 3 times higher than that of the thiobarbituric acid method (0.0037 g/l) for sialic acid determination. The enzymatic-kinetic assay correlates well with the latter method (r = 0.973). There was no measurable interference by EDTA (less than 3 mg/l), bilirubin (less than 160 mumol/l), haemolysis (haemoglobin less than 5.25 g/l), or triglycerides (less than 3.50 g/l). The reference range (2.5th to 97.5th percentile) for sialic acid in sera of adults is 0.51 to 0.84 g/l (median 0.63 g/l) with no significant sex-related differences. The reference range for sialic acid in spontaneous urine samples is 0.01-0.13 g/l.
唾液酸的酶法测定作为一种终点法最近已商业化推出,通过分别开发动力学方法的初速率和峰速率改进方法,有了显著改进,并适用于离心分析仪(Cobas Bio)。该检测在3分钟内完成,样品体积为3微升,总反应体积减少到153微升。因此,每次检测的成本降低到原来方法的约10%。批内和日间不精密度(均值 = 0.81 g/l)分别为2%和3.9%;误差在 +1%至 -2%之间。检测限为0.012 g/l,即比用于唾液酸测定的硫代巴比妥酸法(0.0037 g/l)高约3倍。酶动力学检测与后一种方法相关性良好(r = 0.973)。EDTA(低于3 mg/l)、胆红素(低于160 μmol/l)、溶血(血红蛋白低于5.25 g/l)或甘油三酯(低于3.50 g/l)均无明显干扰。成年人血清中唾液酸的参考范围(第2.5至97.5百分位数)为0.51至0.84 g/l(中位数0.63 g/l),无明显性别差异。自发性尿液样本中唾液酸的参考范围为0.01 - 0.13 g/l。