Yi Kyu-Ho, Park Soo Yeon, Wan Jovian, Song Jong Keun, Kim Jin-Hyun, Koppert Erik, Kim Jae Young
Department of Oral Biology, Human Identification Research Institute, BK21 FOUR Project, Division in Anatomy and Developmental Biology, Yonsei University College of Dentistry, Seoul, Republic of Korea.
You and I Clinic, Seoul, Republic of Korea.
JPRAS Open. 2025 Jul 18;45:473-479. doi: 10.1016/j.jpra.2025.07.011. eCollection 2025 Sep.
The efficacy of thread lifting procedures for facial rejuvenation is significantly influenced by thread fixation strength and tissue stabilization capabilities. This study presents a quantitative biomechanical comparison between bidirectional threads (epiticon® BI, JETEMA Co., Ltd.) combined with multidirectional (epiticon® ORIGINAL, JETEMA Co., Ltd.) threads versus double bidirectional threads.
Ex vivo testing was conducted on fresh cadaver tissue specimens ( = 30 per group). Two experimental configurations-bidirectional plus multidirectional and double bidirectional-were tested using standardized insertion techniques with l-type cannulas. Fixation strength was measured using calibrated tensile testing equipment.
The bidirectional and multidirectional combination demonstrated significantly higher mean fixation strength (52.1 ±2.13 N) compared to double bidirectional (47.35 ± 2.19 N), representing an ≈10 % increase ( = 0.032). Qualitative assessment revealed more uniform tissue engagement patterns with the bidirectional and multidirectional combination, while the double bidirectional combination showed greater tissue concentration at the threads' central portions.
The strategic combination of bidirectional with multidirectional threads provides superior fixation strength and more uniform tissue engagement compared to using bidirectional threads exclusively. These findings have direct clinical implications for optimizing thread selection in facial rejuvenation procedures, potentially enhancing the longevity and naturalness of aesthetic outcomes.
面部年轻化埋线提升手术的效果受线体固定强度和组织稳定能力的显著影响。本研究对双向线(epiticon® BI,JETEMA株式会社)联合多向线(epiticon® ORIGINAL,JETEMA株式会社)与双股双向线进行了定量生物力学比较。
对新鲜尸体组织标本进行离体测试(每组 = 30个)。使用标准化的插入技术,采用l型套管针,对双向线加单向线和双股双向线这两种实验配置进行测试。使用校准后的拉伸测试设备测量固定强度。
双向线和多向线联合使用时的平均固定强度(52.1 ± 2.13 N)显著高于双股双向线(47.35 ± 2.19 N),增加了约10%(P = 0.032)。定性评估显示,双向线和多向线联合使用时组织嵌入模式更均匀,而双股双向线组合在缝线中央部分显示出更大的组织集中。
与单纯使用双向线相比,双向线与多向线的策略性组合提供了更高的固定强度和更均匀的组织嵌入。这些发现对优化面部年轻化手术中的缝线选择具有直接的临床意义,可能会提高美学效果的持久性和自然度。