Besch W, Woltanski K P, Fischer U, Kohnert K D, Ziegler M
Exp Clin Endocrinol. 1985 Dec;86(2):146-50. doi: 10.1055/s-0029-1210480.
A sensitive radioimmunoassay (RIA) for canine C-peptide (CCP) was established using synthetic CCP, a specific antiserum, and rabbit anti-guinea pig serum. Radioiodination was performed according to a modified chloramine-T method. Tracer preparations were used for long as 6 weeks after iodination. The standard curve ranges from 0.028 to 3.0 nmol/l. The intra-assay coefficient of variation (CV) was 3-5% and the inter-assay CV was 6-9% in the optimal range between 0.3 and 0.8 nmol/l. The average recovery of CCP added to plasma samples was 100.6% (n = 9). Canine insulin, porcine proinsulin, bovine proinsulin, and human C-peptide exhibited no cross-reactivity. The mean fasting plasma CCP concentration was 0.089 +/- 0.021 nmol/l in normal dogs and -0.005 +/- 0.007 nmol/l (mean +/- SEM) in diabetic dogs, respectively.