Cohen G, DiBianca F A
J Comput Assist Tomogr. 1979 Apr;3(2):189-95. doi: 10.1097/00004728-197904000-00008.
Limiting-perception measurements of hole patterns in computed tomography reconstructions of a low-contrast phantom are used to generate contrast -- detail -- dose diagrams for various contrasts, hole diameters, dose, and system modulation transfer functions (MTF). These curves are then divided into MTF-dominated, transition, and noise-dominated regions. The low-contrast perceptibility and image noise are parameterized and compared. A limiting-perception factor is difined and is shown to be nearly independent of contrast, hole diameter, dose, and MTF in the noise-dominated region.