Lewellen T K, Nelp W B, Murano R, Palmer H E, Hinn G M
Phys Med Biol. 1979 Jan;24(1):124-35. doi: 10.1088/0031-9155/24/1/010.
A technique to measure total body calcium using the 40Ca(n, alpha)37 Ar reaction has been developed. The technique is based on collecting 37Ar exhaled in the breath following a 10 mrad uniform total body irradiation by 14 MeV neutrons. The 37Ar in the exhaled breath is extracted by selective absorption and its radioactivity is measured inside a low-background proportional detector. The facilities developed include an activation facility providing a +/-2.7% activation uniformity, a closed circuit rebreathing and gas collection system, and a gas purification and counting system. The technique provides a precision of +/-2.4% as determined by repetitive measurements of human volunteers and has an accuracy for determination of total body calcium in grams of +/-5%.