High-resolution determination of the R-R interval.

作者信息

Law H F, Epstein R A, Epstein M A

出版信息

Am J Physiol. 1979 Jun;236(6):H894-8. doi: 10.1152/ajpheart.1979.236.6.H894.

Abstract

A preprocessor is described that determines the time interval between successive R waves of the electrocardiogram with an error of less than 0.2 ms. It accomplishes this by sepearting in hardware the function of determining the approximate location of the R wave from the function of localizing with great accuracy the peak of the wave form. With this device it is possible to detect abnormal heart rate patterns characterized by exceedingly small beat-to-beat variability. The approach may be useful for other bioelectric signals.

摘要

描述了一种预处理器,它能确定心电图连续R波之间的时间间隔,误差小于0.2毫秒。它通过在硬件中将确定R波大致位置的功能与极其精确地定位波形峰值的功能分开来实现这一点。使用该设备可以检测以逐搏变异性极小为特征的异常心率模式。该方法可能对其他生物电信号也有用。