Spellacy W N, Buhi W C, Cruz A C, Gelman S R, Kellner K R, Birk S A
Am J Obstet Gynecol. 1979 Jul 1;134(5):528-31. doi: 10.1016/0002-9378(79)90834-2.
Seventy-eight amniotic fluid samples were obtained by transabdominal amniocentesis and were analyzed for their L/S ratio and optical density at 400 and 650 nm. The L/S ratio was considered to be mature if the values were greater than 2.0. The optical density of the fluids decreased with prolonged refrigeration. With freshly centrifuged samples, the OD650 reading of greater than 0.15 gave the best correlation with the mature L/S ratio. There were 3.8% false positive results and 14.1% false negative results. The study confirms that the OD650 test on fresh amniotic fluid is a rapid and inexpensive way to determine fetal maturity. All samples with values of less than 0.15 must, however, also be tested for their L/S ratio since some of the infants with these values may also be mature.
通过经腹羊膜穿刺术获取了78份羊水样本,并对其卵磷脂/鞘磷脂(L/S)比值以及在400和650纳米处的光密度进行了分析。如果L/S比值大于2.0,则认为其成熟。随着冷藏时间延长,羊水的光密度降低。对于新鲜离心的样本,OD650读数大于0.15与成熟的L/S比值具有最佳相关性。假阳性结果为3.8%,假阴性结果为14.1%。该研究证实,对新鲜羊水进行OD650检测是确定胎儿成熟度的一种快速且经济的方法。然而,所有OD650值小于0.15的样本也必须检测其L/S比值,因为一些具有这些值的婴儿也可能已经成熟。