St. Jude Medical cardiac valve prosthesis: in vitro studies.

作者信息

Emery R W, Nicoloff D M

出版信息

J Thorac Cardiovasc Surg. 1979 Aug;78(2):269-76.

PMID:459535
Abstract

The St. Jude Medical cardiac valve prosthesis is an all-pyrolitic carbon, bileaflet, low profile, central flow device. Prior to clinical use on intensive program of in vitro and in vivo testing was undertaken. Simulated radiographic studies determined that the tungsten impregnation needed to make the valve leaflets radiopaque was 0.013 inch. Steady-state flow and pulse duplicator studies demonstrated a high ratio of flow orifice to tissue anulus diameter with low transvalvular pressure gradients. Regurgitation was found to be acceptable. Flow visualization revealed central flow through all three flow orifices with minimal turbulence and vorticeal current formation. In vitro comparison with other available cardiac valve devices of various sizes demonstrated superior hemodynamic performance of the St. Jude Medical cardiac valve.

摘要

圣犹达医疗心脏瓣膜假体是一种全热解碳、双叶、低轮廓、中心血流装置。在临床使用之前,进行了深入的体外和体内测试项目。模拟放射学研究确定,使瓣膜叶不透射线所需的钨浸渍厚度为0.013英寸。稳态血流和脉搏复制器研究表明,血流孔与组织瓣环直径的比例很高,跨瓣膜压力梯度较低。发现反流情况可以接受。血流可视化显示,通过所有三个血流孔的血流呈中心性,湍流和涡流形成极少。与其他各种尺寸的现有心脏瓣膜装置进行的体外比较表明,圣犹达医疗心脏瓣膜具有卓越的血液动力学性能。