Cameron H U, Harris W R
Arch Orthop Trauma Surg (1978). 1979 Apr 30;93(4):313-4. doi: 10.1007/BF00450232.
A femoral neck fracture following surface replacement of the hip has been studied and five potential causes of failure were noted. These consisted of; removal of a hip nail during surgery, use of a too small femoral cup, initial reaming in varus, notching of the femoral neck, and failure to completely seat the cup.