The Parekh-Jung method for determination of inorganic phosphorus in serum was applied to its determination in urine. Accuracy is good. The mean percentage recovery of added phosphorus was 100.2%. Forty analyses of a sample gave a mean value of 46.71 mg/100 ml +/- 0.76 standard deviation (between-batch precision). The relative usefulness of the Parekh-Jung method compared with those other methods is discussed.

摘要

将用于测定血清中无机磷的帕雷克-荣氏方法应用于尿液中无机磷的测定。准确度良好。添加磷的平均回收率为100.2%。对一个样本进行40次分析得出平均值为46.71毫克/100毫升±0.76标准偏差(批间精密度)。文中讨论了帕雷克-荣氏方法与其他方法相比的相对实用性。