基于区域化学质量控制数据的额外日常精度估计。

Additional day-to-day precision estimates based on regional chemistry quality control data.

作者信息

DiSilvio T V

出版信息

Ann Clin Lab Sci. 1979 Nov-Dec;9(6):474-8.

PMID:518009
Abstract

State-of-the-art precision values are presented for the following serum constituents: aldolase (EC 4.1.2.13), alpha-hydroxybutyrate dehydrogenase (EC 1.1.1.30), cholinesterase (EC 3.1.1.8), cortisol, gamma glutamyl transferase (EC 2.3.2.2), haptoglobin, immunoglobulins, lactic acid, leucine aminopeptidase (EC 3.4.1.1), total lipids, osmolality, protein fractions, T3 uptake, thyroxine and vitamin B12. Precision estimates are based on values reported for four lyophilized serum pools analyzed by participants in the Pennsylvania Association of Clinical Pathologists regional quality control program for clinical chemistry, during 1976, 1977 and 1978. Use of the upper limit of the "most common range" of precision (that range including the 75 percent most precise laboratories) as a warning level for trouble-shooting is advocated.

摘要

给出了以下血清成分的最新精密度值

醛缩酶(EC 4.1.2.13)、α-羟丁酸脱氢酶(EC 1.1.1.30)、胆碱酯酶(EC 3.1.1.8)、皮质醇、γ-谷氨酰转移酶(EC 2.3.2.2)、触珠蛋白、免疫球蛋白、乳酸、亮氨酸氨肽酶(EC 3.4.1.1)、总脂质、渗透压、蛋白质组分、T3摄取、甲状腺素和维生素B12。精密度评估基于1976年、1977年和1978年宾夕法尼亚临床病理学家协会临床化学区域质量控制项目参与者分析的四个冻干血清池报告的值。提倡将精密度“最常见范围”的上限(该范围包括75%最精密的实验室)用作故障排除的警告水平。