Suppr超能文献

具有通道时分的生物遥测系统中的同步

Synchronization in biotelemetry systems with time division of channels.

作者信息

Fraden J

出版信息

Biotelem Patient Monit. 1979;6(4):192-9.

PMID:526573
Abstract

In order to improve synchronization in biotelemetry systems, a relative pulse-width selector was designed. It employs a sawtooth generator, sample-and-hold and analogue comparator. The use of this selector creates a possibility of reducing the increase in the synchronizing pulse with respect to the channel pulses and eliminating tuning the transmitter's modulator and receiver's selector to each other. Two different selectors are described: a simple one for two different pulse widths and a more complicated one with a counter for any setting of pulse width in the system.

摘要

为了提高生物遥测系统中的同步性,设计了一种相对脉宽选择器。它采用锯齿波发生器、采样保持电路和模拟比较器。使用这种选择器可以减少同步脉冲相对于通道脉冲的增加,并消除对发射机调制器和接收机选择器进行相互调谐的需求。文中描述了两种不同的选择器:一种是用于两种不同脉宽的简单选择器,另一种是带有计数器的更复杂选择器,可用于系统中任何脉宽设置。