The ATREC system is designed to carry out at sixty times real time the quantitative evaluation of arrhythmias in 24 hour magnetic tape recordings. A preprocessing unit eliminates the low frequency waves (P and T) and artefacts, and furnishes the computer (Mitra 15/35, 16 K words of core) with information on R-R interval, QRS duration, regular or irregular rhythm, bradycardia or pauses, transient to sustained tachycardia, and ventricular fibrillation. Extrasystoles are analyzed in terms of enlargement, number, coupling interval, bi or trigeminy, doublets or salvos. Trends of mean and extreme cardiac rate, and R-R interval histograms, are also available. The system is also available for ECG real time monitoring of patients in coronary care units. The computer can analyze 8 ECG tracings simultaneously. The ATREC system provides an important aid in the quantification of arrhythmias and improves the evaluation of antiarrhythmic drugs efficacy.