Ahmadi B, Akbarnia B A, Ghobadi F, Ganjavian M S, Nasseri D
Clin Orthop Relat Res. 1979 Nov-Dec(145):150-3.
Three different distraction devices were investigated for the treatment of leg-length discrepancy caused by poliomyelitis in 144 patients. The average leg-length discrepancy was 5.2 cm and lengthening obtained 4.8 cm. For tibial leg-lengthening. Anderson's apparatus was used in 50 patients; an external fixator device (developed in Shafa Rehabilitation Hospital) was used in 40 patients; and the Wagner apparatus was used in 51 patients. The incidence of major complications was low in all 3 groups. The overall failure rate was 4%. There were no significant differences in the final results among the 3 groups. However, preference is given to the Wagner device because of its practical advantages.
对144例因小儿麻痹症导致腿长不等的患者,研究了三种不同的牵引装置用于治疗。平均腿长差异为5.2厘米,延长了4.8厘米。对于胫骨延长,50例患者使用了安德森装置;40例患者使用了一种外固定装置(由沙法康复医院研发);51例患者使用了瓦格纳装置。所有3组的主要并发症发生率都很低。总体失败率为4%。3组的最终结果没有显著差异。然而,由于其实际优势,更倾向于使用瓦格纳装置。