Suppr超能文献

植入式起搏器即将发生故障的检测。

The detection of impending failure in implanted pacemakers.

作者信息

Davies J G, Siddons H

出版信息

Thorax. 1969 Jan;24(1):74-7. doi: 10.1136/thx.24.1.74.

Abstract

As implantable pacemakers last a variable time, it is desirable to be able to foretell when each individual pacemaker is nearing the end of its life. A test which detects a number of different impending electronic failures including battery depletion is described. The test requires only standard laboratory equipment and has been used for four years as a three-monthly procedure in an out-patient clinic. In 17 of 19 consecutive patients the test has enabled a change of pacemaker to be carried out electively. rather than after an arbitrary period or as an emergency when pacing has failed.

摘要

由于植入式心脏起搏器的使用寿命各不相同,因此能够预测每个起搏器何时接近使用寿命结束是很有必要的。本文描述了一种可检测包括电池耗尽在内的多种不同即将发生的电子故障的测试方法。该测试仅需标准实验室设备,已在门诊作为每三个月进行一次的程序使用了四年。在连续19例患者中的17例中,该测试使得能够择期更换起搏器,而不是在任意时间段之后或在起搏失败时作为紧急情况进行更换。