Suppr超能文献

一枚斯塔尔-爱德华兹主动脉瓣人工瓣膜的布片导致右冠状动脉阻塞。

Right coronary artery occlusion by cloth from a Starr-Edwards aortic valve prosthesis.

作者信息

Angelini G D, Kulatilake E N, Armistead S H

出版信息

Thorac Cardiovasc Surg. 1984 Dec;32(6):379-80. doi: 10.1055/s-2007-1023428.

Abstract

We report a case of sudden onset of angina 11 years after implantation of a No. 8 Model 2300 Starr-Edwards cloth-covered aortic valve prosthesis for aortic stenosis and insufficiency. At operation the cloth covering one strut of the metal cage was tenuously attached to the ring of the prosthesis, with its free end completely occluding the right coronary artery ostium. The valve was replaced and the obstructing foreign body removed, re-establishing the patency of the right coronary artery.

摘要

我们报告一例在植入8号2300型斯塔尔-爱德华兹布覆盖式主动脉瓣人工瓣膜治疗主动脉瓣狭窄和关闭不全11年后突然发生心绞痛的病例。手术时,金属笼架一根支柱上的布与人工瓣膜环的附着很薄弱,其游离端完全阻塞了右冠状动脉口。更换了瓣膜并取出了阻塞性异物,恢复了右冠状动脉的通畅。