Hubl W, Taubert H, Freymann E, Meissner D, Stahl F, Dörner G
Exp Clin Endocrinol. 1984 Jul;84(1):63-70. doi: 10.1055/s-0029-1210367.
A sensitive enzyme immunoassay for cortisol in plasma and saliva without extraction has been developed. An antibody was raised in rabbits against cortisol-3-CMO-BSA and compared with anti-cortisol-21-HS-HSA. The enzyme label was a covalently linked cortisol-horseradish peroxidase conjugate. Solvent extraction was avoided by use of heat denaturation of transcortin and dilution of plasma 1:300. This assay had a limit of detection of 3 fmol/tube and satisfied the standard criteria of accuracy and precision. The test requires only a simple photometer. The normal values of cortisol in plasma and saliva before and after ACTH, dexamethasone or metyrapone application were determined. Saliva, being easily collected by a noninvasive, stress-free technique is an alternative to plasma. The cortisol concentrations were significantly increased in plasma and saliva of patients with Cushing's syndrome and decreased in patients with Addison's disease.
已开发出一种用于血浆和唾液中皮质醇的灵敏酶免疫测定法,无需提取。用抗皮质醇-3-CMO-BSA在兔体内制备抗体,并与抗皮质醇-21-HS-HSA进行比较。酶标记物是一种共价连接的皮质醇-辣根过氧化物酶共轭物。通过使用皮质素结合球蛋白的热变性和将血浆稀释1:300避免了溶剂提取。该测定法的检测限为3 fmol/管,满足准确性和精密度的标准标准。该测试仅需要一台简单的光度计。测定了应用促肾上腺皮质激素、地塞米松或甲吡酮前后血浆和唾液中皮质醇的正常值。唾液可通过无创、无压力技术轻松收集,是血浆的替代物。库欣综合征患者的血浆和唾液中皮质醇浓度显著升高,而艾迪生病患者的皮质醇浓度降低。