Narahara K A, Mena I, Maublant J C, Brizendine M, Thompson C J, Criley J M
Cathet Cardiovasc Diagn. 1984;10(4):395-401. doi: 10.1002/ccd.1810100415.
Gold-195m has a halflife of 30.5 sec. Its ultrashort halflife and low radiation exposure (240 mrads to the kidney from a 20-mCi injection) make it an ideal agent for sequential first-transit studies of cardiac function. The reproducibility of ejection fraction and wall motion determinations were assessed from two gold-195m first-pass angiograms performed 3 min apart in 38 subjects. The accuracy of these evaluations was assessed by performing another first-transit study using technetium-99m. The correlation of the ejection fraction between the two gold-195m angiograms was 0.95 (standard error of the estimate (SEE) = 0.049) over an ejection fraction range of 20-78. The correlation of the ejection fraction using technetium-99m versus gold-195m was 0.95; SEE = 0.050. Only six minor discrepancies were noted when wall motion was evaluated in the gold-195m angiograms and compared with the technetium-99m study. We conclude that gold-195m can provide repeated, accurate, and reproducible ejection fraction and wall motion determinations with conventional imaging equipment at a substantially lower radiation exposure than with technetium-99m.
金 - 195m的半衰期为30.5秒。其超短半衰期和低辐射暴露(20毫居里注射剂量对肾脏的辐射为240毫拉德)使其成为用于心脏功能连续首次通过研究的理想试剂。在38名受试者中,间隔3分钟进行两次金 - 195m首次通过血管造影,评估射血分数和壁运动测定的可重复性。通过使用锝 - 99m进行另一项首次通过研究来评估这些评估的准确性。在20 - 78的射血分数范围内,两次金 - 195m血管造影之间的射血分数相关性为0.95(估计标准误差(SEE)= 0.049)。锝 - 99m与金 - 195m的射血分数相关性为0.95;SEE = 0.050。在金 - 195m血管造影中评估壁运动并与锝 - 99m研究进行比较时,仅发现六个小差异。我们得出结论,与锝 - 99m相比,金 - 195m可以使用传统成像设备以低得多的辐射暴露提供重复、准确且可重复的射血分数和壁运动测定。